Download Manual - Can-Am Instruments Ltd. .PDF
9104723 Downloads
6.73 MB .pdf
To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual. A screen will display 1 Point Manual Complete and the slope (XX. ELEC Standard electrode is not performing within the required specifications. Keeping the process electrode and salt bridge moist will avoid slow response when the sensor is placed back in operation. This can be verified by monitoring the sensor temperature value for a stable temperature measurement. 7 ORP Calibration The manufacturer offers a one point manual calibration for ORP. In normal operation the touch screen displays the measured values for the sensors selected. Products may not be returned without authorization from Hach Company. If an entry is incorrect, repeat the entry with the correct values. FLASH FAILURE System flash memory write has failed. SELECT BUFFER Select the buffer type (standard 4, 7, 10 or DIN 19267) from the displayed choices. See Control Drawing 58600-78 in the sc100 Controller Manual, Cat. SELECT SENSOR Choose from the displayed sensor type (pH or ORP). A screen will display 1 Point Manual Complete and the slope (XX. CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury. Keep the button pressed to change the numbers/characters continuously. TEMP OUT RANGE Measured temperature exceeds the expected value range. If surface deposits cannot be removed by detergent solution cleaning, use muriatic acid (or other dilute acid) to dissolve them. If the entry is outside the working range, a correction to the entry is made automatically. From the Main Menu, select SENSOR SETUP and confirm