Nanodrop Instructions
General info
UV-Vis Spectrophotometer
Only needs 1-2 µL sample
Assesses quality and quantity
Wireless data transfer- will it reach 5th floor from Nanodrop on 4th floor? Probably easiest to use USB flash drive
View data using Nanodrop One/Oneᶜ Viewer software
Important instructions:
Only wipe upper and lower pedestals with a clean, dry, lint-free wipe. Do NOT use a paper towel
Do NOT spray or squirt any liquid onto the NanoDrop
Only micropipettes should be used to dispense liquid onto the lower pedestal
When measuring samples IMMEDIATELY click Measure after lowering the sampling arm
Wipe pedestals clean ASAP after taking sample measurements so the sample doesn’t dry on the pedestals
Use a fresh pipette tip and fresh drop of sample for each measurement
Always clean the pedestals with dH₂O and a lint-free wipe at the very beginning and very end of taking measurements
When you go down to Dr. Chang’s lab on the 4th floor bring:
Samples
Micropipette
Pipette tips
Gloves
Also don’t forget to introduce yourself and ask Dr. Chang if it’s okay to use the Nanodrop
Measuring DNA and RNA
Clean pedestal surfaces prior to starting
Cleaning
Use dH₂O and a lint-free laboratory wipe to clean upper and lower pedestals, and then dry with kimwipe
Do NOT spray or squirt water onto NanoDrop
If there is a dried sample stuck on the pedestals use a 3 µL drop of HCl followed by 3 µL of dH₂O to clean
Reconditioning (DO NOT HAVE TO DO)
Use the applicator in the reconditioning kit, take a pin-head sized amount of NanoDrop Pedestal Reconditioning Compound (PR-1)
Apply a thin even layer of PR-1 to the flat part of the lower and upper pedestals
Wait 30 seconds for the PR-1 to dry
Fold a lint-free laboratory wipe into quarters and wipe PR-1 off until no more dark PR-1 compound residue shows up on the wipe
Take a blank measurement using the same buffer as used in the sample
Pipette 1- 2 µL of blank buffer (usually MBG H₂O) onto lower pedestal
Lower sampling arm
Click Blank
Wipe pedestals with dry, lint-free wipe
Do NOT use samples or blank buffers containing hydrofluoric acid (HF)
dH₂O and TE are common blank buffers used for nucleic acids
Gently vortex sample
Note nucleic acid sample must be purified prior to measurement
Pipette 1-2 µL of the sample onto the lower measurement pedestal
1 µL is usually sufficient but 2 µL is better if the sample has reduced surface tension properties
0.5 µL can be used if the sample has 10 mm equivalent absorbance values of 3.0 or higher (>150 ng/ µL dsDNA)
Do not use samples with dye!
Lower sampling arm (has upper pedestal)
Click Measure IMMEDIATELY after lowering sampling arm
Make sure drop of liquid (sample column) was touching both the lower pedestal and the upper pedestal of the sampling arm
When done with measurement wipe both the upper and lower pedestals with a lint-free laboratory wipe
Use a fresh pipette tip and a fresh drop of sample everytime. Failure to do so will result in increased concentration measurements due to evaporation.
If measuring many samples take a new blank every 30 minutes
When done clean with dH₂O and a lint-free laboratory wipe
Troubleshooting unusual spectra:
Negative absorbance values: the pedestal is dirty or there was something wrong with the blank➝ clean the pedestals and take a new blank measurement
Ragged appearance (lines not smooth): bad blank➝ clean pedestal and redo blank
Jagged appearance (very rough lines almost like a zigzag): broken sample column➝ clean and recondition both pedestals and try increasing sample volume (no more than 2 µL)
High 230 nm absorbance value relative to sample: contaminated sample➝ may need to redo sample
Instructions for interpreting the NanoDrop results for DNA are found here:
https://docs.google.com/document/d/1hsAzy2MWxDj2EdsFWOlJUFh-A3jtAZ4Cf6RnZnpHLRA/edit
Measuring Proteins
Clean and recondition pedestal surfaces prior to starting
Cleaning
Use dH₂O and a lint-free laboratory wipe to clean upper and lower pedestals
Do NOT spray or squirt water onto NanoDrop
If there is a dried sample stuck on the pedestals use a 3 µL drop of HCl followed by 3 µL of dH₂O to clean
Reconditioning
Use the applicator in the reconditioning kit, take a pin-head sized amount of NanoDrop Pedestal Reconditioning Compound (PR-1)
Apply a thin even layer of PR-1 to the flat part of the lower and upper pedestals
Wait 30 seconds for the PR-1 to dry
Fold a lint-free laboratory wipe into quarters and wipe PR-1 off until no more dark PR-1 compound residue shows up on the wipe
Take a blank measurement using the same buffer as used in the sample
Many common buffers are not suitable for measurement
To check if the buffer is suitable:
Pipette 2 µL of dH₂O onto lower pedestal
Lower sampling arm
Click Blank
Wipe pedestals with a dry, lint-free wipe
Pipette 2 µL of blank buffer onto lower pedestal
Lower sampling arm
Click Measure
Wipe pedestals with a lint-free laboratory wipe
Analyze spectra- should be a straight line that vaies no more than 0.4 mm from baseline
If buffer is compatible continue
Pipette 2 µL of a new sample of blank buffer onto lower pedestal
Lower sampling arm
Click Blank
If it is already known that the buffer is suitable:
Pipette 2 µL of blank buffer onto lower pedestal
Lower sampling arm
Click Blank
Enter sample name in Sample ID box and choose appropriate sample type
Enter extinction coefficient and/or molecular weight if asked (see guide)
Lightly vortex sample
Note protein sample must be purified prior to measurement
Pipette 2 µL of the sample onto the lower pedestal
Lower sampling arm
Click Measure IMMEDIATELY after lowering sampling arm
Make sure drop of liquid (sample column) was touching both the lower pedestal and the upper pedestal of the sampling arm
When done with measurement wipe both the upper and lower pedestals with a lint-free laboratory wipe
Use a fresh pipette tip and a fresh drop of sample everytime. Failure to do so will result in increased concentration measurements due to evaporation.
If measuring many samples take a new blank every 30 minutes
When done clean with dH₂O and a lint-free laboratory wipe
Troubleshooting unusual spectra:
Negative absorbance values: the pedestal is dirty or there was something wrong with the blank➝ clean the pedestals and take a new blank measurement
Ragged appearance (lines not smooth): bad blank➝ clean pedestal and redo blank
Jagged appearance (very rough lines almost like a zigzag): broken sample column➝ clean and recondition both pedestals and try increasing sample volume (no more than 2 µL)
Jagged appearance at the top of the 280 nm peak: sample too concentrated➝ dilute sample and redo all measurements