Electrophysiology FAQs, Tips, and Troubleshooting
Adjustment Knob of Micromanipulators
Problem: Sometimes when using the electrophysiology set up by the hallway (not by the windows), one of the channel 2 adjustment knobs will slip.
To fix this, loosen the adjustment knob that holds the black tube that comes from the blue box holding the micropipette (after you put the oocyte in the well and you are ready to impale). Push the channel 2 holder down towards the oocyte to get it close to the solution. Do not impale this way. Once it is close, tighten the knob. Now you can use the micromanipulators and they should not slip.
Setting Up an Experiment
How do I know the glass pipette is good?
Before filling the glass pipette, check the tip with the microscope on the other side of the lab between the pipette puller and the dissecting scope used for oocytes. Ensure that the tip is neither too large nor too small, and make sure that it has not been broken. If you have measured the pipette tip, include the measurement in your experiment analysis near the oocyte information (RNA and concentration, etc.).
How do I know how much to fill the glass pipettes?
Place the empty glass pipette into the pipette holder and mark with your finger where the pipette emerges from the holder. Using this as a guide, fill the glass pipette with KCl up to this point, ensuring that no KCl makes its way into the pipette holder when the pipette is inserted. If the liquid gets into the pipette holder, rinse the pipette holder and use a different holder for the experiment.
Why isn’t there a valid channel 1/ channel 2 reading?
Before the tip of the pipette enters the FR solution, the voltage should read “1” or “-1.” If the voltage reading does not change when the tip of the pipette is placed in FR, there is likely a contact issue. First, ensure that there is adequate KCl in the pipette but not so much that it has entered the pipette holder. Also ensure that there is sufficient contact between the silver wire (shown below) and the pipette holder. If the issue persists, check to see if a holder that works on one of the channels will also work on the other channel. If the pipette holder fails to produce a reading on the other channel, it is likely an issue with the pipette holder. If the channel fails to give a reading regardless of which pipette holder is used, try to adjust the knob so that a voltage reading is displayed.
Also check to make sure the electrode well is filled with KCl and you have placed the salt bridge. If either of these steps were skipped, you will have an invalid reading.
Is it okay to make all the necessary solutions ahead of time so I don’t have to worry about making them during the experiment?
Usually, it’s best to make just the first few solutions ahead of time. That way, if the oocyte isn’t expressing well or something goes wrong early in the experiment, we won’t have to waste or store unused solutions. If you still want to prepare somewhat, you can measure out I+I for the solutions ahead of time and just add reagents when the time comes to mix the solutions.
Preparing Oocytes for Electrophysiology
How do I choose a good oocyte?
First, check the dates on the oocytes. Make sure that they are neither too old nor too recently injected with RNA (this allows time for sufficient expression of the protein of interest). Look at the oocytes under the dissecting scope. Choose an oocyte that is a decent size, intact (does not appear to be disintegrating or falling apart) and has good contrast between the yellow and brown hemispheres, if possible.
How do I place the oocyte in the chamber?
The goal of this process is to orient the oocyte with the yellow side up so that it can be impaled. This process requires a bit of trial and error. Use the pipette to drop the oocyte in the chamber a bit upstream of the middle well and allow the oocyte to roll with the flow of frog ringer and come to rest in the well. If the oocyte is not properly oriented, gently use the pipette to remove the oocyte from the well and try again.
How should the oocyte be impaled?
Have someone experienced train you on this, as it is a more complicated part of electrophysiology experiment set up. Overall, the micropipette tips should impale the oocyte at approximately 45 degrees on either side, or about halfway up the oocyte. You should place the tips just at the surface of the membrane to dimple it, but do not use the micromanipulators to actually pierce the membrane. Instead, tap one manipulator at a time, checking the voltage for that channel. When the voltage jumps very negative, you know that you have impaled. A picture of the locations of the micropipettes on the oocyte is below for your reference.
How do I know when to start the experiment?
Before starting the experiment, the voltage readings should have stabilized and be fairly consistent (usually less than 10 difference in the readings) across both channels. You should also “play” a measurement (this is different from recording a measurement; when you “play” a reading, it is not saved for the experiment file) and ensure that the resulting red line has very little slope/is almost flat and is smooth. When you are confident in this baseline reading, zero the stopwatch and begin the experiment.
Running an Experiment
What should I do if the flow stops during the experiment?
Try to keep an eye on the solution so that this does not occur. If you’re using a very small amount of solution and you are worried that the cup will run dry, you can tilt the cup to pool the solution at one end and take an early last measurement if necessary to avoid using up the last of the solution and stopping the flow. However, if it does occur, make a note that the flow stopped (and when during the experiment, if possible) and include this information on your experiment analysis. You may try to restart the flow, but be very gentle and slow when pushing solution into the tube, as too much force can cause the oocyte to be pushed to the side in the chamber, possibly disrupting the contact with the pipettes or severely damaging the oocyte. If the oocyte becomes severely damaged, abort the experiment.
One of the pipettes slipped out of the oocyte during the experiment. Now what?
Usually, when this happens the voltage of only one of the channels drops and reads near zero while the other channel is much more negative, unlike the voltage changes that affect both channels in response to a solution change. Turn on the microscope light, check the oocyte in the chamber using the microscope, and attempt to reimpale. Reposition the pipette corresponding to the affected channel on the oocyte so that a slight dimple forms and the voltage reads slightly negative. Very gently tap one of the adjustment knobs on for the pipette. If the voltage begins to read negative (and ultimately a similar value to the other channel), you have likely successfully reimpaled the oocyte and can continue the experiment.
How to tell, and what to do if I get a block?
Blocks occur during an experiment when the end of the channel 1 or channel 2 pipettes become blocked by the cellular contents of an oocyte (ooooh-site). You can tell this is happening when you notice a random (messy) drop in the line when viewing the raw data (Figure 1). To get yourself out of this mess, you will want to spam current through your electrodes. To do this, you will flip the current switch from record to clamp (like you would when doing an actual experiment) and press the black play button (pictured below #2) until you see the line straighten and return to normal. (You won't always be able to get the block out)
Ion Substitution Experiments
I switched the solution to sodium iodide and now the voltage measurement is positive. Is something wrong?
Although unusual, this actually isn’t a problem, so you don’t need to fix anything. Continue with the experiment.
Why can’t MTSET and MTSES solutions be made ahead of time?
When making the solution for ET or ES modification, don’t make it until right before you need to switch the solution. It degrades very quickly (within minutes) out of the freezer. To quickly thaw the aliquots after removing them from the freezer, use a 1000µL pipette to transfer the pre-measured I+I for the solution into the microcentrifuge tube and gently pipette up and down to mix before pipetting the contents into the cup. Also make sure you wash the tube containing the ET and ES multiple times with the micropipette to make sure you get as many molecules out of the tube as possible.
Analyzing Experiments
How do tell if an experiment is good?
Compare your analyzed experiment to representative experiments and prior experiments. Using this information, see if your experiment followed prior trends and expectations for the type of RNA and type of experiment conducted. It helps to look at both bad and good experiments and distinguish the differences between them
Connection Issues
What if I've messed with the pipette holders and nothing is working?
to be updated
What is Background Noise?
If your baseline looks like what is pictured here, you have background noise.
to be updated