Many solutions can be coloured.
You may already be familiar with some coloured aqueous solutions, for example:
The colour we see is the colour of the light that is transmitted through the solution.
"White light" is part of the electromagnetic spectrum and is made up of all the colours of the rainbow; red, orange, yellow, green, blue, indigo and violet (ROYGBIV).
If we shine white light through an aqueous solution of copper sulfate, the solution appears to be blue because it is transmitting the blue wavelengths of white light and absorbing other wavelengths such as red so that very little of these other wavelengths are transmitted through the solution to your eye.
The colour of the solution is due to the wavelengths of light that are not absorbed as "white light" passes through the solution.
Now, if we dilute the original copper sulfate solution by adding more water, the blue colour becomes less intense, it is a lighter blue colour because the solution is not absorbing as much of the other wavelengths as before:
If we were given an aqueous solution of copper sulfate of unknown concentration, solution (A), we might be able to take a guess at its concentration by looking at its colour:
If our solution is blue, we choose a light colour that is complementary to blue, that is, we choose the colour that is opposite this colour in the colour wheel, in this case we choose an orange light to shine on our blue solution.
The Beer-Lambert law relates the attenuation of light to the properties of the material through which the light is traveling. We will take a brief look at the Beer-Lambert Law and explains the use of the terms absorbance and molar absorptivity relating to UV-visible absorption spectrometry.
For each wavelength of light passing through the spectrometer, the intensity of the light passing through the reference cell is measured. This is usually referred to as Io - that's I for Intensity.
The intensity of the light passing through the sample cell is also measured for that wavelength - given the symbol, I. If I is less than Io, then the sample has absorbed some of the light (neglecting reflection of light off the cuvette surface). A simple bit of math is then done in the computer to convert this into something called the absorbance of the sample - given the symbol, A. The absorbance of a transition depends on two external assumptions.
Assumption one relates the absorbance to concentration and can be expressed as
The absorbance (A) is defined via the incident intensity Io and transmitted intensity I by
Assumption two can be expressed as
Combining Equations 1 and 3
This proportionality can be converted into an equality by including a proportionality constant (ϵ).
This formula is the common form of the Beer-Lambert Law, although it can be also written in terms of intensities:
The constant ϵ is called molar absorptivity or molar extinction coefficient and is a measure of the probability of the electronic transition. On most of the diagrams you will come across, the absorbance ranges from 0 to 1, but it can go higher than that. An absorbance of 0 at some wavelength means that no light of that particular wavelength has been absorbed. The intensities of the sample and reference beam are both the same, so the ratio Io/I is 1 and the Log10 of 1 is zero.
You will find that various different symbols are given for some of the terms in the equation - particularly for the concentration and the solution length.
The Greek letter epsilon in these equations is called the molar absorptivity - or sometimes the molar absorption coefficient. The larger the molar absorptivity, the more probable the electronic transition. In uv spectroscopy, the concentration of the sample solution is measured in mol L-1 and the length of the light path in cm. Thus, given that absorbance is unitless, the units of molar absorptivity are L mol-1 cm-1. However, since the units of molar absorptivity is always the above, it is customarily reported without units.
The proportion of the light absorbed will depend on how many molecules it interacts with. Suppose you have got a strongly colored organic dye. If it is in a reasonably concentrated solution, it will have a very high absorbance because there are lots of molecules to interact with the light. However, in an incredibly dilute solution, it may be very difficult to see that it is colored at all. The absorbance is going to be very low. Suppose then that you wanted to compare this dye with a different compound. Unless you took care to make allowance for the concentration, you couldn't make any sensible comparisons about which one absorbed the most light.
Suppose this time that you had a very dilute solution of the dye in a cube-shaped container so that the light traveled 1 cm through it. The absorbance is not likely to be very high. On the other hand, suppose you passed the light through a tube 100 cm long containing the same solution. More light would be absorbed because it interacts with more molecules. Again, if you want to draw sensible comparisons between solutions, you have to allow for the length of the solution the light is passing through. Both concentration and solution length are allowed for in the Beer-Lambert Law.
Next we plot these results on a graph and draw a "line of best fit" through the data points:
Now we measure the absorbance of our aqueous solution of copper sulfate of unknown concentration.
Example: Absorbance of unknown solution = 0.950
We plot the point (x) for this unknown solution on our calibration curve:
Reading off the graph we see that an absorbance of 0.950 corresponds to a concentration of 0.34 mol L-1
Alternatively, if we know the mathematical equation for the straight line of the calibration curve, we can use that to determine the concentration of our unknown solution.
On the calibration graph we were given the expression:
Abs = 2.78 × [CuSO4(aq)]
which is the mathematical equation for the calibration curve.
Substitute the value for the absorbance of our unknown solution into the equation:
0.950 = 2.78 × [CuSO4(aq)]
Divide both sides of the equation by 2.78:
(0.950 ÷ 2.78) = (2.78 ÷ 2.78) × [CuSO4(aq)]
Solve the equation to find the concentration of our unknown solution:
0.3417 mol L-1 = [CuSO4(aq)]
However, we are only justified in using 2 significant figures to express the concentration of our unknown solution, that is:
[CuSO4(aq)] = 0.34 mol L-1
In summary, to determine the concentration of a coloured solution using colorimetry:
The Problem: Chris the Chemist needs to determine the concentration of a deep blue copper sulfate solution.
First Chris transfers 25.00 mL of the solution by pipette into a 100.00 mL volumetric flask and makes the solution up to the mark with distilled water.
Next, Chris prepares copper sulfate solutions of known concentration and measures their absorbance in order to establish a calibration curve as shown below:
Then Chris measures the absorbance of the diluted solution and finds that it is 0.506
Determine the concentration of the original copper sulfate solution.
Step 1: Plot the absorbance for the diluted solution on the calibration curve to determine its concentration.
Step 2: Determine the concentration of the undiluted original solution:
ciVi = cfVf
ci = [CuSO4(aq)(original)] = ? mol L-1
Vi = volume of CuSO4(aq)(original) = 25.00 mL
cf = [CuSO4(aq)(diluted)] = ? mol L-1 (from calibration curve)
Vf = volume of CuSO4(aq)(diluted) = 100.00 mL
[CuSO4(aq)(diluted)] = 0.18 mol L-1
ciVi = cfVf
ci = [CuSO4(aq)(original)] = ? mol L-1
Vi = volume of CuSO4(aq)(original) = 25.00 mL = 25.00 mL ÷ 1000 mL/L = 0.02500 L
cf = [CuSO4(aq)(diluted)] = 0.18 mol L-1 (from calibration curve)
Vf = volume of CuSO4(aq)(diluted) = 100.00 mL = 100.00 mL ÷ 1000 mL/L = 0.10000 L
ci × 0.02500 L = 0.18 mol L-1 × 0.10000 L
ci × 0.02500 L = 0.018 mol
ci × 0.02500 L ÷ 0.02500 L = 0.018 mol ÷ 0.02500 L
ci = 0.72 mol L-1 (only justified in 2 significant figures)
There are two basic approaches as to how you vary the transition metal ion – ligand ratio and the results illustrated in the diagram below.
Method (1) The mole ratio method keeping Mn+ constant and gradually increasing the number of moles of ligand X from zero to a large molar excess.
For dilute solutions of copper(II) ions and ammonia the graph gradient change occurs at 4.0 moles of NH3per mol of Cu2+ for the reaction ...
The manganate(VII) ion, MnO4–, e.g. in potassium manganate(VII) solution, is a brilliant purple colour and its concentration in very dilute solution can be measured by using colorimetry i.e. by comparing the absorbance of the solution versus a calibration graph of known concentrations of the manganate(VII) ion.