mAs: 21.5
kVp: 90
EI : 941
DI: 0.4 (good)
Changes to be made: 35–40-degree oblique position. CR slightly too low. Should be at the level of T5-T6 or 1 inch inferior to sternal angle. I believe it may have been taken AP instead of PA. Or the image was not flipped in post processing. Entire esophagus is demonstrated. Evaluation criteria states that the esophagus should be midway between the spine and heart. No marker present.
mAs: 24.4
kVp: 90
EI : 349
DI: 0.6 (good)
Changes to be made: CR to T5- T6, possibliy a touch too high. Entire esophagus demonstrated. Esophagus midway between the heart and spine and the arms are not superimposing the esophagus. Marker visible, good exposure factors. Collimation could have been brought in more.
mAs: 12.8
kVp: 90
EI : 1026
DI: 1.6
Changes to be made: Patient in a 30-60 degree oblique (phantom placed at a 45 degree angle). CR to L1. Entire stomach and duodenum are demonstracted and the fundus would be barium filled in this position. Optimal exposure factors and marker is present but in the anatomy. Collimation could have been bought in more.
mAs: 10.8
kVp: 90
EI : 799
DI: 0.5 (good)
Changes to be made: CR to L1. Entire stomach and duodenum demonstrated. Body and pylorus should be barium filled with the air in the fundus. Optimal exposure factors with a marker present.