mAs: BU: 449.2mAs/1560ms before exposure. 16.2 mAs after exposure. 55.4 ms after exposure.
kVp: 120
EI: 441 (very low)
Changes to be made: Exposure was too low and not enough contrast. mAs needs to be raised. Patient is not completely perpendicular and even with IR.
mAs: BU: 499 mAs 156ms before exposure. Post exposure 8.2 mAs.
kVp: 115
EI: 611 (low)
Changes to be made: Low exposure, mAs should have been raised or kVp for more contrast. Slightly tilted and could have been centered on the IR better.
mAs: 3.20 before exposure. 3.41 mAs after exposure.
kVp: 110
EI: 1875 (high)
Changes need to be made: The sternal end of the clavicles were uneven. Patient was rotated and tilted to one side. The EI was high and high contrast. mAs should be lowered. Cannot see marker on image.
mAs: 4.00
kVp: 120
EI: 2,135 (high)
Changes to be made: Patient/IR was off center. Tilted and rotated, the sternal ends of the clavicle are uneven. Shoulders are not rotated and covering the tops of the lungs. The contrast/mAs is too high and needs to be lower with more grays.