mAs: 4.00 mAs/ 3.97 after
kVp: 55
mA; 322.7
Seconds:12.3 ms
EI : 1626
Changes to be made: mAs decreased, kvp increased in the 70-75 kVp range. Soft tissue around the elbow not present. Closer collimation.
mAs: 2.5 mAs
kVp: 60
S number : 118
Changes to be made: S number is within range but it looks as though there isnt enough density on the image. Placement good but needs closer collimation.
mAs: 4.0/ 3.97 after
kVp: 55
mA; 323.9 mA
Seconds: 12.2 ms
EI : 1547
Changes to be made: Collimation could have been brought in more. mAs reduced, kVp increased. Data drop out in the soft tissue of the elbow. Marker present.
mAs: 4.00
kVp: 55
mA; 322.4
Seconds: 12.3
EI : 1619
Changes to be made: Marker present, superimposed radius and ulna. Closer collimation.
mAs: 2.60
kVp: 78
mA; 320
Seconds: 8.0 ms
EI : 1869
Changes to be made: Collimation needs to be extened to include more of the elbow joint. Orentiation needs to be changed post processing.
mAs: 4.00 before/3.97 after
kVp: 55
mA; 322.4
Seconds: 12.3 ms
EI : 1781
Changes to be made: Artifacts present due to needing glove boxes to hold phantom. Elbow and wrist joints included. Radial and ulta superimposed. Could have moved CR down a little. Reduce mAs increase kVp.