mAs: 8
kVp: 76
EI : 662
DI: 3.4
Changes to be made: Atlas, Axis and dens not seen. CR correct positioning. Upper incisors not superimpsed over the base of the skull. No rotation present. Marker present but in the anatomy. Exposure level is too high. Collimation could have been brought in more.
mAs: 8
kVp: 75
EI : 839
DI: 1.8
Changes to be made: CR 15 degrees cephalad to lower thyroid cartilage. C3 to the area of T2 demonstrated. No rotation present but a slight tilt to the right. Intervertebral spaces pen. Mandible is cut off and base of the skull is not super imposing C1-C2. Too close of collimation. Marker present.
mAs: 8
kVp: 76
EI : 1705
DI: 7.5
Changes to be made: C1-C7/T1 demonstrated. Rami of mandible not superimposed over C1/C2. No rotation present. Collimation needed to be brought in more to not include the jaw and skull. Marker present.
mAs: 20
kVp: 75
EI : 1333
DI: 1.9
Changes to be made: CR 3-4 inches below the juglar notch. Collimation needs to be brought in more on 4 sides. No rotation or tilt present. CR is positioned slightly too far to the left. Vertebral bodies well penetrated. Marker present.
mAs: 32
kVp: 78
EI : 269
DI: -4.1
Changes to be made: CR perpendicular to T7. Collimation needs to be brought in more and post processing completed. Marker present but in the anatomy. Range of T1-L1. Intervertebral disk spaces open. No rotation present. Orthostatic breathing needed to blur lung markings.
mAs: 20
kVp: 80
EI : 32
Changes to be made: Collimation needs to be brought in more to include less of the L-spine. No rotation present. Vertebral bodies well penetrated but could have increased the mAs/kVp. CR is too low.
mAs: 32
kVp: 80
EI : 45
Changes to be made: Upper thoracic posterior portions of the vertebral bodies are cut off. Range of T1-L1 included. CR too anterior on phantom. Marker present.
mAs: 32
kVp: 80
EI : 49
Changes to be made: Phantom in true lateral position. Right shoulder needed to be dropped. CR to T1. Tube angled 5 degrees cephalic. Humeral heads slightly separated.