Electron Tomography - Taking a tilt series using EMMEMU

[MODES]

[TEM LOW DOSE]

Enable joystick Z (for tilt series):

[COMPUSTAGE]

[COMPUCTRL]

[JOYSTICK Z]

§ Ready

Without inserting a holder do flatfields

Make sure that the vacuum < 17

Bean on by rotating filament knob slowly.

flatfields / background subtraction:

confirm beam is centered and spread the beam to get meter=2.06 s

raise screen; cover view field with mat

Repeat process below for each camera format used:

Format

> select desired format (typically,2k x 2k)

OK

Flatfields

Exposure (uncorrected)

adjust exposure to desired mean (typically 2000)

Acquisition:

Exposures: 4 (min) – 8 (max)

all boxes unchecked

Create Flatfield

Check and Save:

Exposure (corrected)

should show featureless grainy image

Store Flatfield

Close

Beam off

Load a sample onto a high tilt holder and insert it.

Wait until vacuum gets down to 16

Bean on

[SEARCH]

Format

> 0: 2048X2048, 1X, …

OK

take test exposure to update format

adjust exposure time for appropriate mean.

align [SEARCH] and [EXPOSURE] modes

for [EXPOSURE] mode:

D (diffraction mode on)

use eyepiece focus needle for correlation

[SEARCH]

> Spot size: spot 3

use highest Spot for which you can see detail

find a unique “landmark” near region of interest and adjust to focus

[EXPOSURE]

> Spot size: spot 6

D (diffraction mode on)

Adjust to focus and center on [SEARCH] area using multi-function knobs

[SEARCH]

adjust to focus

Exposure to test

[RESET DEFOCUS]

higher mag = lower defocus

-4um for 25,000X

-10um for 12,000X

to check precise defocus

Power

Power [>] Fit PCTF

increase defocus to fit rings to zeros

Apply

Adjust defocus as necessary

Exposure to test

Format

> 1: 1024X1024, 2X, …

OK

Set exposure to 100 ms and take test exposure to update format

adjust exposure time for appropriate mean if necessary.

Keep the screen up

Positions

Show positions

SCANNING, EXPOSURE, TRAKING tab, check magnification

Calibration …

Update TEM

Calibrate Gonio X/Y

Calibrate tilt axis

Calibrate communication

Calibrate image shift

Calibrate beam shift

follow instructions

Close

Show positions

SCANNING tab

Beam shift x:0

Beam shift y:0

Image shift x: 0

Image shift y: 0

Check the spot size

Apply

Exposure, tracking, and Search tab

Check sopt size

Apply and close

> SCANNING

Goto position

Update Position

> ALL

Copy position to

√ Lock positions

Show positions

SCANNING tab

Defocus: 0

Offset defocus: 0

EXPOSURE tab

Format:

> 0: 2048X2048, 1X, …

Exp. time (ms): 50-100% of standard exposure time

check mean, modify as needed

Defocus: 0

Offset defocus: 0

FOCUS tab

Defocus: -5000

Offset defocus: -5000

TRACKING tab (for tilt series)

Exp. time (ms) adjust for same mean as EXPOSURE mode

Defocus: -5000

Offset defocus: -5000

SEARCH tab

Format:

> 1: 1024X1024, 2X, …

Magnification: ~½ mag for EXPOSURE mode

Exp. time (ms): lowest possible exposure time to yield detail

Defocus: -20000

Offset defocus: -20000

Apply and Close

Close

test beam centering:

use desired magnification

focus beam and center with shift x/y knobs

raise and lower screen to check for beam shift

repeat calibrations if necessary

move to an area close to your object

eccentricity correction

[EXPOSURE]

§ D (diffraction mode on)

y coordinate must be within 100 of origin

correct for eccentricity near region of interest

same y coord as region of interest

[COMPUSTAGE]

[A-WOBBLER] on

adjust Z so field doesn’t shift

[A-WOBBLER] off

[EXPOSURE]

Goniometer Tilt

tilt to ±70°

adjust Z so area stays in field for all tilt angles

[SEARCH]

Exposure

check focus and eccentricity. compare with previous exposure

[EXPOSURE]

· D (diffraction mode on)

relocate to region of interest

v Goniometer Tilt

tilt to -70° or 70°

raise screen and cover view screen with the mat

[SEARCH]

Positions

Center detail

Update Position

> ALL

Copy position to

Areas

tracking should be outside of exposure area along tilt axis

search area should include exposure area

Tilt series

change directory and file name as needed

Start angle: -70 or as needed

End angle: 70 or as needed

√ Exposure every

√ Tracking every

√ Search every

√ Focus every (if not cryo)

√ Angles by Saxton scheme with 0° increment 2 or 3 deg

√ Fully automatic tracking

√ Display focusing steps

√ Display tracking steps

START

¤ Add entry before: 0

Entry to add:

Position:

¤ Exposure

Angle: current angle

Do edit

¤ Search

Angle: current angle

Do edit

adjust meter to 2.06

cover view field w/ mat

lift screen

Start series

to manually recenter during tilt series acquisition

press spacebar at XCF screen

Continue semi-automatic

Center detail

recenter using tracking image

Continue full-automatic

Convert .dat format to .tif: and generate save_txt file

user date

make sure spe ang. spe dir. mean are checked.

close

Open series

¤ EM-FORMAT

CONTINUE >>

Save series

Same as saving regular images