Located in room SC2202.
Below the specifications is the calendar to schedule appointments, and below that is the calendar showing all already scheduled reservations:
You must be logged into a @georgiasouthern.edu or Gmail account to reserve instrument time.
The Shimadzu LCMS-8030 is a powerful Triple Quadrupole Liquid Chromatography-Mass Spectrometry (LC-MS/MS) system used for highly sensitive and accurate quantification and confirmation of trace-level analytes in various matrices.
Types of Samples and Applications:
This system is particularly well-suited for the analysis of samples in complex matrices and is employed in various fields, including:
Biological Specimens: Detecting and quantifying drugs and metabolites in biological samples.
Environmental Monitoring: Analyzing environmental contaminants and pesticides in various samples.
Food and Beverage Analysis: Screening for pesticides and other contaminants in food and beverages.
Pharmaceutical Testing: Identifying and quantifying genotoxic impurities, as well as performing Bioanalytical (BA) and Bioequivalence (BE) studies.
Biotechnology, Clinical, Diagnostic, and Research Laboratories: Used for various applications requiring sensitive and quantitative analysis.
Capabilities of the LCMS-8030 Triple Quad UV-VIS:
The LCMS-8030 boasts several features and capabilities that make it a powerful analytical tool:
Triple Quadrupole Mass Spectrometry (TQ-MS): This technique is ideal for sensitive quantification and confirmation of target analytes.
Ultra-Fast MRM Transitions: Offers rapid analysis with dwell times as low as 1 msec and pause times of 1 msec, enabling high-speed multi-analyte detection (up to 555 channels per second).
Ultra-Fast Polarity Switching: Achieves rapid polarity switching in just 15 msec, allowing for comprehensive analysis of both positive and negative ions.
UFsweeper® Technology: A patented technology that accelerates ions out of the collision cell for efficient collision-induced dissociation (CID) and ultra-fast ion transport, minimizing sensitivity losses and crosstalk.
High-Speed Scanning: Features a high scanning rate of up to 15,000 u/sec, allowing for full spectrum scans within a series of MRM measurements.
Robust and High-Sensitivity Detection: The heated ion source handles challenging samples and offers high-sensitivity detection using ESI, APCI, or a dual-probe ionization interface.
Easy Maintenance: Streamlined maintenance procedures, including quick and straightforward cleaning of the heated desolvation capillary, ensure greater system uptime and usability.
Seamless Integration with UHPLC Systems: When coupled with Shimadzu's Nexera UHPLC, the LCMS-8030 provides a unified platform for enhanced qualitative and quantitative analysis, increased productivity, and accelerated workflows.
Note: The UV-VIS detection component enhances the system's capabilities by providing orthogonal detection for compounds with UV-Vis absorbance. Triple quadrupole mass spectrometry primarily provides information based on the mass-to-charge ratio and fragmentation patterns of ions, while UV-Vis detection is based on the absorption of light at specific wavelengths.