Evaluation of Light Curing Units and Dentists’ Knowledge About Photo Polymerization Techniques in Sulaimani Governmental Dental Clinics

* Mohammed A. Mahmood, ** Bestoon M. Faraj

* Kurdistan Board for Medical Specialties, Iraq. 

** Kurdistan Board for Medical Specialties and Department of Conservative Dentistry, College of Dentistry, University of Sulaimani, Sulaimani, Iraq. 

Submitted: 17/03/2019; Accepted: 30/05/2019; Published 01/06/2019

DOI: https://doi.org/10.17656/sdj.10088

Abstract

Objective: To measure the output intensity of light curing units (LCU) in governmental dental clinics of the Sulaimani governorate. To evaluate practitioners’ knowledge on light cure application.

Methods: Ninety-four dental units and eighty-eight dentists in the Sulaimani governorate were included in this research. Output intensity and diameter of tips of the dental LCU devices were measured. Two survey questionnaires were filled, one for the LCU and its maintenance, the other for evaluating the knowledge, attitude, and practice of the dentists. 

Results: Output power intensity results ranged from 50-2000W, with an average of 993W. The output intensity of 78% of the tested devices was found to be acceptable regarding the manufacturer instruction of the radiometers used (every intensity below the given range that given in the Result section is regarded as unusable or non-acceptable). The range of age in service of the devices was 1-12 years with a mean of 4.7 years. The average restorations performed in a week by a device was 17.8. Dentists had poor knowledge, attitude, and practice towards the maintenance of the LCU devices. 51% of the dentists were quite satisfied with the devices. 48.9% of the dentists reported that they don’t have an idea of LCU intensity.

Conclusions: Devices should be checked regularly for output intensity, light bulb efficiency, and composite build up. Dentists express poor knowledge, maintenance, and attitude toward LCUs. Educational programs are advisable and recommended.

                                                                                                                                                                                                                                                                                                                                             

Keywords: Light-cure intensity, Dentists knowledge.                                                                                                                                                                                                                                                               Full Article - PDF

                                                                                                                                                                                                                                                                                                                                                      

References:

1, Sakaguchi R, Douglas W, Peters M. Curing light performance and polymerization of composite restorative materials. J Dent. 1992;20(3):183-8. 

2. Mitton B, Wilson N. Practice maintenance: The use and maintenance of visible light activating units in general practice. Br Dent J. 2001;191(2):82. 

3. Shortall A, Harrington E, Wilson H. Light curing unit effectiveness assessed by dental radiometers. J Dent. 1995;23(4):227-32. 

4. Madhusudhana K, Swathi T, Suneelkumar C, Lavanya A. A clinical survey of the output intensity of light curing units in dental offices across the Nellore urban area. SRM J Res Dent Sci. 2016;7(2):64. 

5. Santini A. Current status of visible light activation units and the curing of light-activated resin-based composite materials. Dent. update. 2010;37(4):214-27. 

6. Rueggeberg F, Caughman WF, Curtis J. Effect of light intensity and exposure duration on cure of resin composite. Oper. Dent. 1994;19:26-. 

7. Berry T, Barghi N, Godwin J, Hunter K. Measurement of intensity of curing light units in dental offices. J Dent Res. 1992;71:161.

8. Hegde V, Jadhav S, Aher GB. A clinical survey of the output intensity of 200 light curing units in dental offices across Maharashtra. J conserv dent. 2009;12(3):105. 

9. Ab Rahman A, Husein A, Ahmed HMA, Mohamad D, Bakar WZW, Farea M, et al. A survey on light intensity outputs of QTH, cabled and cordless LED light curing units. Arch Orofac Sci. 2014;9:85-90. 

10. Pilo R, Oelgiesser D, Cardash H. A survey of output intensity and potential for depth of cure among light-curing units in clinical use. J Dent. 1999;27(3):235-41.  

11. Santini A, Turner S. General dental practitioners' knowledge of polymerization of resin-based composite restorations and light curing unit technology. Br Dent J. 2011;211(6):E13. 

12. Alsuliman SA. Knowledge of Dental Practitioners about Light Curing Units in Saudi Arabia. Int J Dent & Oral Heal. 2018;4(9):127-9. 

13. Nuray Tüloğlu, Sezin Özer, Emine Şen Tunç, Sevgi Canbaz, Şule Bayrak. Knowledge And Attitudes of Dental Clinicians Regarding LightCuring Units In Northern Turkey. cdr. 2016;40(1):26-34. 

14. Burgess JO, Walker RS, Porche CJ, Rappold AJ. Light curing--an update. Compendium of continuing education in dentistry (Jamesburg, NJ: 1995). 2002;23(10):889-92. 

15. Hansen EK, Asmussen E. Correlation between depth of cure and surface hardness of a lightactivated resin. Eur J Oral Sci. 1993;101(1):62-4. 

16. Yogesh Sahu VK, Aditi Jain, Pooja Deshmukh, Deepti Jain, Shruti Sharma. To evaluate the functional efficacy of various light curing units in dental clinics across Chhattisgarh state. IJCE. 2017;2(1):1-5. 

17. Maghaireh GA, Alzraikat H, Taha NA. Assessing the irradiance delivered from light-curing units in private dental offices in Jordan. J Am Dent Assoc. 2013;144(8):922-7.  

18. Pires JAF, Cvitko E, Denehy GE, Swift Jr EJ. Effects of curing tip distance on light intensity and composite resin microhardness. Quintessence Int. 1993;24. 

19. Oberländer H, Friedl K-H, Schmalz G, Hiller K-A, Kopp A. Clinical performance of polyacidmodified resin restorations using” softstartpolymerization”. Clin Oral Investig. 1999;3(2):55-61. 

20. Kwon YH, Kwon T-Y, Ong JL, Kim K-H. Lightpolymerized compomers: coefficient of thermal expansion and microhardness. J Prosthet Dent. 2002;88(4):396-401. 

21. Nomoto R, Uchida K, Hirasawa T. Effect of light intensity on polymerization of light-cured composite resins. Dent Mater J. 1994;13(2):198-205.

  

 © The Authors, published by University of Sulaimani, College of Dentistry

This work is licensed under a Creative Commons Attribution 4.0 International License.