Experiment Details
Dimensions = 1
Isotope labeling required = 13N
Magnetization transfer = 1H - 13C
Pulse program: hC.cp (basic cross polarization experiment)
Parameters:
F1, C and F2, H
p1 : C 90 at pl1(acqt0-calculation & -DC90)
p3 : H 90 at pl2
p15 : HC CP at sp41 (f1,C) and sp40 (f2,H)
pl1 : C pulse power
pl2 : H pulse power
pl12 : H dec power
sp40 : H CP power
sp41 : C CP power
d1 : recycle delay; 1 to 5 times T1
pcpd2 : pulse length in decoupling sequence cpdprg2
cpdprg2 : spinal64, swftppm, tppm etc. decoupling program
spnam40 : H ramp use e.g. ramp.10070 for variable amplitude CP
spnam41 : C ramp use e.g. square.100 for square pulse CP
ZGOPTNS : -DC90 : optimize C 90 power
-Dlacq : aq is longer than 50 ms or blank
Comment on key optimization:
ZGOPTNS = '-DC90' for C 90 cal
Example spectrum of GB1: Recorded at MAS 17857 Hz in a 3.2 mm rotor.