Here is some important information you need to know before using the NMR instruments.
EinsteinNMR USER ACCOUNT: Each user must have an NMR instrument account and undergo a brief training before using instruments for first time. To become an NMR User, please see: New User Info
RESOURCE ACCESS: the door for Ullmann B-08 is kept on bypass while the resource is staffed during the daytime. A code is needed for access at all other times; users will receive the code when they get an account and are trained.
EMERGENCY CONTACT: the cell number of staff is posted by each instrument. If you run into a problem with the instrument such as the software crashes or becomes unresponsive, please contact the staff - please do not attempt to fix any computer hardware or NMR hardware without consulting with staff first.
PRECAUTIONS AND RESOURCE RULES:
No Food or drinks are allowed in the Magnet Lab in Ullmann B8
Trained NMR Users are allowed to use instruments on their own as long as they have reserved the instrument and have an account on the instrument they are using
NO ONE EXCEPT FOR TRAINED NMR USERS ARE ALLOWED WITHIN THE CHAINED OFF AREA OF EACH MAGNET. Guests of users MUST stay outside the chained off area.
Magnetic Field Hazards:
Magnets will exert large attractive forces on equipment or other magnetic objects when brought close. The force may become large enough to move the equipment uncontrollably towards the NMR magnet. Small pieces of metallic subjects (wrenches, screwdrivers...) may become projectiles. Large equipment (gas cylinders) can cause bodies or limbs to become trapped between the equipment and the magnet. Keep in mind the following:
The closer to the magnet, the larger the force.
The larger the mass, the larger the force.
REMEMBER: Superconducting magnets are ALWAYS on!!!
Rules and Info for working near magnets
Do NOT bring any metallic object within 10 feet of any magnet. Assume all metallic objects are ferrormagnetic and will be attracted to the magnets, unless verified by NMR staff.
Do NOT bring compressed gas cylinders into the NMR laboratories without NMR staff supervision.
NEVER put any object into the magnet, except the NMR tube and holder.
For low-temperature NMR experiments, use the non-magnetic nitrogen dewar that is available. Do not use a wrench to disconnect the quick connector on the dewar.
Medical Implants:
The operation of electronic,electrical or mechanical medical implants, such as cardiac pacemakers, biostimulators, and neurostimulators may be affected or even stopped in the presence of either static or changing magnetic fields. Medical implants such as aneurysm clips, surgical clips or protheses may contain ferromagnetic materials and therefore would be subjected to strong attractive forces near the magnet. This could result in injury or death. Additionally, in the vicinity of rapidly changing fields (pulsed gradient fields), eddy currents may be induced in the implant resulting in heat generation.
Persons with these types of implants MUST remain outside the NMR laboratories until more extensive safety training is provided by NMR staff.
Cryogen Hazards:
Cryogens such as liquid nitrogen (LN2) and liquid helium that are present in the magnet cryostat and portable dewars may pose several dangers, including asphyxiation, frostbite and chemical explosions.
A) When a magnet quenches, or suddenly becomes non-superconducting, large amounts of liquid cryogens are quickly vaporized. Here is a video of a magnet quenching. Due to their large expansion ratios (nitrogen 695:1, helium 760:1), these gases can quickly displace all the oxygen in the NMR room and cause asphyxiation. Effects from oxygen deficiency become noticeable at levels below ~18% and sudden death may occur at ~6% oxygen content by volume.
B) Direct contact with cryogenic substances in liquid or vapor form can produce “cold burns” on the skin similar to conventional burns. The temperature of liquid helium is -269 C and of liquid nitrogen is -196 C.
C) Cryogenic fluids with a boiling point below that of liquid oxygen are able to condense oxygen from the atmosphere. Repeated replenishment of the system can thereby cause oxygen to accumulate. Violent reactions, e.g. rapid combustion or explosion, may occur if the materials, which make contact with the oxygen, are combustible.
Rules and Info regarding cryogens:
If you observe a sudden exhaust of gas from a magnet (and NMR staff are not performing a cryogen fill), EXIT the NMR laboratory IMMEDIATELY.
NMR staff periodically must replenish the magnet’s cryogens. During a fill, keep away from the gaseous exhaust from the magnet as frostbite burns may result.
When handling cryogens, wear gloves, googles, and closed toe shoes.
When doing low-temperature NMR work and using the LN2 dewar, beware of liquid splashing and rapid flash off of LN2 when immersing the variable-temperature NMR apparatus into the dewar. This operation must be carried out very slowly.
Always replace the loose-fitting cap onto the LN2 dewar to avoid the condensation of oxygen in the dewar.
NOTE: There is an oxygen meter by the Bruker 600 computer - if it goes off, please evacuate the lab and contact staff as soon as possible.
NMR Tube Safety:
NMR tubes have thin walls and are easily broken. Once broken, they are extremely sharp. When inserting NMR tube into spinner, grap the tube close to the spinner. This will avoid applying a torque that can easily break a tube and, often, drive it into a finger.
Chemical Safety- Hazardous and dangerous materials:
NOTE: ABSOLUTELY NO RADIOACTIVE SAMPLES ARE ALLOWED IN THE MAGNETS - RUNNING SUCH SAMPLES WILL VOID THE SERVICE CONTRACTS WE HAVE ON THE INSTRUMENTS!!!
All users who conduct experiments with hazardous materials, including toxic materials, including pressurized, explosive, or otherwise unusually hazardous materials MUST FIRST INFORM THE EinsteinNMR RESOURCE STAFF and closely follow Einstein Environmental Health and Safety guidelines. Failure to do so may jeopardize the environment and safety of all users and may lead to suspension of facility access privilege and payment of any cleanup cost.
NMR tubes are often broken in and around the NMRs. Please clean up non-hazardous spills and broken glass. For toxic or hazardous materials, do not attempt to clean up yourself. Instead, inform EinsteinNMR Resource staff and contact Einstein Environmental Health and Safety.