This instrument is capable of detecting up to 18 fluorescent parameters, making it ideal for multicolor flow cytometry assays that involve a large panel of markers. It comes with High throughput Sampler Unit for sample acquisition in multi-well plate
Location: Room 531, Basic Medical Science Building, University of Manitoba Bannatyne Campus, 745 Bannatyne Ave, Winnipeg, MB.
Sample Handling
Semi-automatic Single Tube Loading with 12 x 75 mm or 1.2 mL tubes
High Throughput Sampler unit for multi-well plates (standard and high throughput modes)
3 Lasers:
50mW 488nm Solid State blue laser,
40mW 640nm Solid State red laser
50mW 405nm Solid State violet laser.
50mW 561nm Solid State yellow-green laser
20mW 355nm Solid State UV laser
20 parameters:
1-2: Two fluorescent detectors off blue laser
3-5: Three fluorescent detectors off red laser
6-11: Six fluorescent detectors off violet laser
12-15: Four fluorescent detectors off yellow-green laser
16-18: Three fluorescent detectors off UV laser
19: Forward scatter
20: Side scatter
Acquisition/Analysis software: FACSDiva
Note:
How to read longpass filter detection range
e.g., 502LP—wavelenge Longer than 502nm will be allowed to Pass through
How to read bandpass filter detection range
e.g., 530/30 = 530nm ± (30/2)nm = 530nm ± 15nm = from 515nm to 545nm