This powerful instrument, capable of detecting up to 10 parameters and 8 colors simultaneously, is ideal for both clinical and research applications. Its newly designed loader carousel allows for automatic loading of up to forty 12x75mm tubes without operator intervention, reducing hands-on time and improving results reliability.
Location: Room 466, Apotex Center, University of Manitoba Bannatyne Campus, 750 McDermot Ave, Winnipeg, MB.
Sample Handling
Semi-automatic Single Tube Loading with 12 x 75 mm tubes
3 Lasers:
20mW 488nm Solid State blue laser,
17mW 633nm HeNe red laser, and
30mW 405nm Solid State violet laser.
10 parameters:
1-4. Four fluorescent detectors off 488nm,
5-6. Two fluorescent detectors off 633nm,
7-8. Two fluorescent parameters off 405nm,
9. Forward scatter, and
10. Side scatter.
Acquisition/Analysis software: FACSDiva
Note:
How to read longpass filter detection range
e.g., 502LP—wavelength Longer than 502nm will be allowed to Pass through
How to read bandpass filter detection range
e.g., 530/30 = 530nm ± (30/2)nm = 530nm ± 15nm = from 515nm to 545nm