Before starting your procedure, connect the Camera to the Camera Control Unit (CCU) and the Endoscope. Connect the Light Cable to the Light Source and the Camera. Turn on the light and perform White Balance. Focus the Camera.
Turn on the device.
Connect the Camera to Camera Control Unit.
Choose the specialty.
Perform White Balance. It can be perform via CCU. Alternatively, you can contact your Stryker representative to program any button on the Camera Head to perform White Balance.
Connect Camera Head to the Coupler.
To perform White Balance, hold the button until White Balance Complete shows on the Monitor upper corner .
Coupler has 4 components:
Rear Adaptor which threads onto Camera Head. Grasping the rear adapter, screw the coupler onto the camera head (clockwise) until it forms a tight seal.
Focusing Ring which adjust coupler focus.
Scope End which receives the Endoscope.
Endobody Clamp which secure the endoscope to the coupler. To attach the endoscope, depress the Endobody Clamp, insert the endoscope into the Endobody and release the clamp.
It's LED.
Connect Light Cable to the Stryker L11 Light Source.
Connect Light Cable to Laparoscope.
If Stryker 1688 CCU is paired with Stryker L11 LED Light Source, you can turn on or off Auto Light feature from CCU.