Clean autoclave, Make sure the water is out.
Fill the autoclave with 1,5l water
Fill bag (autoclave bag) Don’t completely close it! Leave hole of 2,5 cm !!
Place the grid in the barrel of the autoclave
Lay the bag on the grid. Make sure it does not touch the walls of the barrel.
Screw on het lid. Work in a cross, tighten one side and then the opposite site. screw first all screws loosely before tightening. Don't tighten too much.
Open the air escape valve, it should stand upright.
Turn on the preheating, wait till the safety valve makes noise and the preheat light turns off.
Turn of the preheat
Close the air escape valve, the tip should be pointing down
Turn on the timer
Set the right time on the timer (to do this, turn the button clockwise ten minutes further then the needed time and turn it counterclockwise back to the needed time)
Wait till the timer makes noise.
Turn off the timer
Let it cool down completely (1 hour)
Open the air escape valve
Open the screws
Get the bag out