Make: Anton Paar
Model: Multiwave GO Plus
Facility Status: Working
Category: Microwave Digestion System
Available Equipment/Mode of Use: Microwave-assisted wet digestion for sample preparation.
Facility Management Team and Location: CGC TEC, Council of Scientific and Industrial Research–National Physical Laboratory (CSIR–NPL)
Department: Environmental Sciences & Biomedical Metrology
LabEmailID:tuhin@nplindia.org, tuhinkumarmandal@gmail.com
Facility Location: Centre for Global Change, QSL Building, TEC
Lab Phone No: 01145609437
Facility Features, Working Principle, and Specifications:
Features:
Rapid and uniform heating for efficient sample digestion.
Pressure and temperature monitoring for safety.
Compatible with Teflon bomb digestion tubes.
Working Principle:
Employs microwave energy with direct coupling to heat aqua regia (3:1 HNO₃:HCl) in a closed system, decomposing metals into soluble nitrates for elemental analysis.
Body Specification:
Temperature range: Up to 300°C.
Pressure range: Up to 40 bar.
Capacity: Suitable for multiple samples in batch processing.
Applications:
Digestion of road dust for ICP-OES analysis of elements like Cd, Cr, Ni, Pb.
Preparation of samples for pollution assessment indices (Igeo, CF, PER).
General wet digestion in environmental and chemical research.