I would like to acknowledge the exceptional work of the Iowa neonatology team and the team by El-Khuffash A. et al., that made some of these below calculators possible in the first place. Please go read their full articles by clicking the respective links.
Original description of the Iowa Score in 2020
From the article: "The following equation was used to derive the risk score for each infant: (gestation in weeks × −1.304) + (PDA diameter in mm × 0.781) + (LVO in mL/kg/min × 0.008) + (maximum PDA velocity in m/s × −1.065) + (LV a′ wave in cm/s × −0.470) + 41, where 41 is the constant of the formula. This score ranges between 0 (low risk) and 13 (high risk). The predicted probability for each infant to develop CLD/death was also derived from the model. Infants without a PDA at the time of the scan were assigned a risk of 0 as a severity score cannot be derived from infants without a PDA diameter or maximum PDA velocity measurement."
References:
"WHAT IS A NORMAL LEVEL OF CARDIAC OUTPUT? A ventricular output of 150 to 300 mL/kg/min is considered normal in (pre)term infants without transductal or interatrial shunting. A ventricular output less than 150 mL/kg/min or SVCf less than 40 to 45 mL/kg/min is associated with adverse outcomes. Although it would be preferred to measure cardiac output very accurately in absolute numbers, it might be more useful to categorize the level of cardiac output (low, normal, or high) for the purpose of understanding the underlying pathophysiology and when interpreted in conjunction with blood pressure for the classification of the stage of shock.37 The combination of low cardiac output and normal blood pressure would suggest a compensated shock, whereas low cardiac output and low blood pressure is indicative of an uncompensated shock. In a hyperdynamic shock, one would expect high cardiac output and low blood pressure. The interpretation of simultaneously assessed cardiac output and blood pressure enables an individualized, pathophysiology-based approach toward cardiocirculatory failure in critically ill newborn infants."
Click on the Spreedsheet to open and enter datapoints for calculation of Z-scores. Alternatively, use the Parameterz tool provided above, which includes the option for these Z-scores.
NeoCardioLab is pleased to announce that the Canadian Congenital and Pediatric Cardiology Research Network Z-score calculator is available right above. This tool is based on the landmark publication: Lauzon-Schnittka J, Plante V, Dahdah N, Greenway SC, Drolet C, Wong K, Mackie AS, Mertens L, Cavallé-Garrido T, Penslar J, Wong D, Garrido-García LM, Dallaire F; Canadian Congenital and Pediatric Cardiology Research Network. Z Scores for Pediatric Echocardiography Dimensions Adjusted for Body Size, BMI, and Age. Circ Cardiovasc Imaging. 2025 Aug;18(8):e017944. doi: 10.1161/CIRCIMAGING.124.017944. Epub 2025 Jul 9. PMID: 40631677. This comprehensive study provides updated reference values for pediatric echocardiographic dimensions—including ventricular, atrial, and great vessel measurements—adjusted for body size, BMI, and age, reflecting a large, multicentre Canadian cohort. The calculator allows clinicians and researchers to derive precise Z-scores for cardiac structures across the pediatric age spectrum, integrating this new reference data into daily echocardiographic interpretation. You need to open (click on top right corner) in order to input the data for Z-score calculations. At NeoCardioLab, this is the calculator we recommend for performing normative Z-score calculations in newborns.
NeoCardioLab a le plaisir d’annoncer que le calculateur de scores Z du Réseau canadien de recherche en cardiologie congénitale et pédiatrique est disponible juste au dessus. Ce nouvel outil est basé sur la publication phare suivante : Lauzon-Schnittka J, Plante V, Dahdah N, Greenway SC, Drolet C, Wong K, Mackie AS, Mertens L, Cavallé-Garrido T, Penslar J, Wong D, Garrido-García LM, Dallaire F; Canadian Congenital and Pediatric Cardiology Research Network. Z Scores for Pediatric Echocardiography Dimensions Adjusted for Body Size, BMI, and Age. Circ Cardiovasc Imaging. 2025 Aug;18(8):e017944. doi: 10.1161/CIRCIMAGING.124.017944. Epub 2025 Jul 9. PMID: 40631677. Cette étude complète fournit des valeurs de référence mises à jour pour les dimensions échocardiographiques pédiatriques, incluant les mesures ventriculaires, atriales et des gros vaisseaux, ajustées en fonction de la taille corporelle, de l’IMC et de l’âge, et reflétant une large cohorte canadienne multicentrique. Le calculateur permet aux cliniciens et aux chercheurs d’obtenir des scores Z précis pour les structures cardiaques à travers tout le spectre pédiatrique, intégrant ces nouvelles données de référence dans l’interprétation échocardiographique quotidienne. Vous devez ouvrir (cliquer sur le coin supérieur droit) afin d’entrer les données pour les calculs de scores Z. Au NeoCardioLab, il s’agit du calculateur que nous recommandons pour effectuer les calculs de valeurs Z normatives chez les nouveau-nés.
Calculator based on the fitted (predicted) reference values
Calculator based on the observed values
References:
For the dp/dt of the LV, we use the CW-Doppler of the MR jet. Here it takes 10.56 msec to get from 1 m/s to 3 m/s. Here the dp/dt is 3030 mmHg/sec, well above the 1200 mmHg/sec threshold for LV systolic dysfunction
In comparaison, for the RV - we take the CW-Doppler of the TR jet. Here it takes 15.84 msec to go from 1 m/s to 2 m/s for the RV by the CW-Doppler of the TRJ. It represents 758mmHg/s, which is above ≥ 400 mmHg/s, likely normal.