Trigger
A trigger is what the ventilator senses to give a patient-initiated breath
There are two types of mechanisms: changes in pressure & changes in flow
Pressure-triggering generally requires the patient to drop the end-expiratory pressure by about 1-4
Pressure-triggering is problematic in patients w/COPD, asthma, or other conditions predisposing to dynamic hyperinflation (auto-PEEP) because they will need to generate a higher inspiratory pressure
Ineffective triggering can be recognized on exam as inspiratory efforts that do not lead to a ventilator-initiated breath
Most ventilators also allow the trigger to be initiated by inspiratory flow, usually in the range of 1-6 L/min
If the flow trigger is too low, then it may "auto-cycle" even when the patient is not trying to initiate a breath
Oscillations in ventilator tubing (i.e., water, secretions, cardiac contractions, patient movement) may inadvertently trigger a breath
The solution is to make the trigger less sensitive or switch to pressure triggeringÂ
Flow
Dynamic Hyperinflation