Versatile electroporation system suitable for applications for gene, drug and protein delivery including: mammalian cells, in vivo, ex vivo, in ovo tissue, ex plant, nuclear transfer and CRISPR NEW! Now available with Pulse Switcher for automatic polarity switching!

The ECM 830 is a versatile electroporation system suitable for applications for gene, drug and protein delivery including: CRISPR, mammalian cells, in vivo, ex vivo, in ovo tissue, ex plant and nuclear transfer. 

ECM 830 GENERATOR

The ECM 830 is a square wave pulse generator designed for in vitro and in vivo applications. The versatility of the ECM 830 applications for gene, drug and protein delivery include; mammalian cells, in vivo, ex vivo tissue, zebra fish tissue and embryos, nuclear transfer, embryo manipulation, plant protoplast and basic bacteria and yeast transformations.

The ECM 830 can be used in combination with a wide array of BTX specialty electrodes and accessories to enhance your molecular and drug delivery experiments.

The ECM 830 is intended For Research Use Only.  Not for use in diagnostic, pre-clinical, or clinical procedures.


FEATURES


APPLICATIONS


ADHERENT CELL TRANSFECTION

The ECM 830 generator coupled with the Petri Pulser electrode or the adherent cell electrode allows researchers to avoid the trypsinization of their cell by electroporating adherent cells directly in the dish they are growing in. The Petri Pulser electrode is ideal for 6-well plates and the adherent cell electrodes for 12-well plates.

HIGH THROUGHPUT (HT) ELECTROPORATION (45-0664 and 45-0666)

The High Throughput (HT) 96- and 25-well systems offer the advantage of multi-well electroporation technology for processing multiple samples in seconds. 

This permits a large number of samples to be quickly electroporated using high throughput plate handler technology. Using the HT Multi-Well plates instead of traditional cuvettes, the researcher can transition from a single cuvette to either a 96-well or 25-well electroporation plate by using a plate handler. This increases yields and the number of experiment runs in a single day. Experiments take seconds to run, allowing for quick and efficient optimization of the electrical and biological parameters. Once optimized, samples are rapidly processed increasing yields and saving valuable time and money.

The key to the HT System is the combination of the ECM 830 Generator with the Plate Handler and multi-well plates. The Plate Handler holds the plates and delivers the pulse(s) to the wells using parameters set in the ECM 830 generator.

MONITORING OPTION

The ENHANCER 3000 allows the researcher to monitor and track key electrical parameters used in electroporation applications. The electrical pulse data is captured as both a graphic display of the wave form and electrical output values following each electroporation experiment. 

AUTOMATIC POLARITY SWITCHING

Combine the ECM 830 with the Pulse Switcher to incorporate automatic bipolar switching into electroporation protocols. Polarity switching enhances transfection efficiency by preventing transfectant accumulation at a single electrode and enhancing delivery to the target cells or tissue.

INCLUDED ITEMS

Item #

Product

Included Items

45-2052

ECM 830 Electroporation System Complete, with Safety Dome

ECM 830 Generator, Safety Dome, Cuvettes 1 mm, 2 mm, 4 mm pkg. of 30 (10 each) and Cuvette Rack 660

45-0661

ECM 830 Electroporation System Complete, with Safety Stand

ECM 830 Generator, 630B Safety Stand, Cuvettes 1 mm, 2 mm, 4 mm pkg. of 30 (10 each) and Cuvette Rack 660

45-0662

ECM 830 Generator Only

ECM 830 Generator Only

45-0663

ECM 830 High Throughput Electroporation System, 25-well with HT-100

ECM 830 Generator, 25-Well Plates (4 mm gap, 6X), Plate Seals, HT-100 Plate Handler and a plate adaptor

45-0664

ECM 830 High Throughput Electroporation System, 25-well with HT-200

ECM 830 Generator, Safety Dome, Cuvettes 1 mm, 2 mm, 4 mm pkg. of 30 (10 each), Cuvette Rack 660, 25-Well Plates (4 mm gap, 6X), Plate Seals, HT-200 Plate Handler and a plate adaptor

45-0665

ECM 830 High Throughput Electroporation System, 96-well with HT-100

ECM 830 Generator, 96-Well Plates (2 mm gap, 2X), Plate Seals, HT-100 Plate Handler and a plate adaptor

 45-0666

ECM 830 High Throughput Electroporation System, 96-well with HT-200

ECM 830 Generator, Safety Dome, Cuvettes 1 mm, 2 mm, 4 mm pkg. of 30 (10 each), Cuvette Rack 660, 96-Well Plates (4 mm gap, 2X), Plate Seals, HT-200 Plate Handler and a plate adaptor

45-0667

ECM 830 Electroporation System Complete, for Cuvettes with Monitoring

ECM 830 Generator, 630B Safety Stand, Cuvettes 1 mm, 2 mm, 4 mm pkg. of 30 (10 each), Cuvette Rack 660, ENHANCER 3000 Probe, ENHANCER Interface Box, Oscilloscope and Cables

45-0668

ECM 830 High Throughput Electroporation System with Monitoring

ECM 830 Generator, Safety Dome, Cuvettes 1 mm, 2 mm, 4 mm pkg. of 30 (10 each), Cuvette Rack 660, ENHANCER 3000 Probe, ENHANCER Interface Box, Oscilloscope, Cables, 25-Well Plates (2 mm gap, 6X), HT-200 Plate Handler and a plate adaptor

 45-0669

ECM 830 Pulse Switcher Bipolar Electroporation System

ECM 830 Generator, Pulse Switcher, Power Supply and USB connector cable, Safety Stand, Cuvettes 1 mm, 2 mm, 4 mm (pkg. of 30, 10 each), and Cuvette Rack

SPECIFICATIONS FOR THE ECM 830 GENERATOR

Display Type : 7 inch color touchscreen

Power Source Voltage : 

100 to 240 VAC, 50 to 60 Hz, CAT II Power 500 W (Pulsed), 50 W (Idle)

Fusing : Type 3AG, 5 A, 250 V; Type T (time delay)


Environmental Characteristics 

Intended Use : Indoor use only

Operating Temperature : 10°C to 40°C

Cooling : Convection through metal case

Relative Humidity :60%

Altitude : 2,000 m (operating)


Mechanical Characteristics 

Footprint : 12.75" x 11.25" x 8.5" (W x D x H)

Weight : 16.0 lb (7.3 kg)

 

STANDARD CAPABILITIES

Operational Status : Internal self test upon start-up

Interface : 7 inch Touchscreen User Interface

Input : 110 to 240 VAC

Charge Time : LV <7 s, HV <4 s

Pulse Length : Range 10 μs – 999 μs LV Mode/ 1 μs resolution 1 msec – 999 msec LV Mode/ 1 msec resolution 1 sec – 10 sec LV Mode/ 0.1 sec resolution 10 μs – 600 μs HV Mode/ 1 μs resolution

Voltage Range : 5 – 500 V LV Mode/ 1 V resolution 505 – 3,000 V HV Mode/ 5 V resolution

Multiple Pulsing : 1 – 99

Pulse Interval : 100 msec – 10 sec

Programmability : Storage up to 1000 User Defined Protocols

Safety : Pre-pulse Sample Resistance check, Over-current Pulse Abort, Instrument Arc Protection

Capacitance : 3,775 μF LV, 85 μF HV