Active Delipidation With SmartClear II Pro

Reagents / Equipment Required:

Protocols

Day 0

[Note] Only 400 mL of solution will be used in the device, so 100 mL can be removed from the bottle to be used for this purpose. This 100 mL can be re-used multiple times. 

Day 1

[Note] Use of older membrane types will result in rapid pH drop and reddening of samples. The v2 membranes also have a grid pattern on the membrane (see image below). 

4. Prepare the samples.

5. Load the samples into the device.

6. Prepare the device for clearing.

[Note] If you want further control of temperature, operate the device in Expert Mode. For preservation of RNA for future FISH studies, clearing should be performed at 37°C. To operate at this temperature, you may need to reduce the current from 1500 mA to 1000 or 1200 mA. 


[Note]: Unique Tissue Types

If your sample is liver, we have found that an additional step will result in more transparent samples. Please follow the Liver Clearing Protocol to complete clearing of livers after Step 1 in the protocol above. 

If your sample contains significant connective tissue, adipose tissue, or skin, (such as whole mouse heads) we have found that an additional dehydrating clearing step helps clear these other tissue types without damaging fluorescent proteins. Please follow the fDisco protocol after completing Step 1 in the protocol above. Please note that this step has the potential to harm antigenicity. This step can also be moved to after immunolabeling if you find that to be the case. 


7. Transfer to PBS with 0.02% sodium azide at the end of the experiment.

8. Wash and shut down the device.

At the end of delipidation, the samples will look the same as before clearing and should not have changed size. You will not be able to tell that the sample is done clearing until the index matching stage. See images below.

You have finished Delipidation and may now move on to the next step. If your sample requires labeling, move on to that section. If no labeling is required, you can move on directly to Index Matching.