Testing of the system was unfortunately unable to be completed due to the COVID-19 pandemic. The Discovery Park building closed on March 23rd, 2020 and instruction moved online to protect the health and safety of UNT students and staff. This section discusses the plans for testing procedures. Test matrices for the CO2 humidifying tank and vortex separator are shown and discussed in tables 5 and 6. The instrumentation of the system is shown in figure 13 and is further described in table 5. Figure 14 shows the overall system and how instrumentation is connected to the system. Data Acquisition and analysis is also discussed, which was to be performed with use of the software LabView and Sensirion Evaluation Kit.
Table 1 demonstrates the planned test matrix for initial analysis of the CO2 humidifying tank. Predicted theoretical values are shown which were planned to be tested against through use of instrumentation, LabView software, and Sensirion Evaluation software.
CO2 flow rate for all runs was based off theoretical calculations previously discussed in the analysis section of this paper. It was theoretically predicted that for a temperature of water in the CO2 humidifying tank at 893.3K a CO2 flow rate of 4.9161*10-5 m3/sec at the inlet to the sparger is necessary to remove maximum required levels of H2O from the humidified CO2 stream given to the X-Hab team by NASA. These values were predicted to cause 100% relative humidity to occur at the gas outlet of the CO2 humidifying tank. The CO2 flow rate is to be varied with temperature of water in the tank, as seen in table 4. Varying both CO2 flow rate and the temperature of water in the tank will allow conditions to be verified for 100% relative humidity to occur at the gas outlet of the humidifying tank.
Table 1: Test Matrix (Broken Up) for CO2 Humidifying Tank
Table 6 demonstrates the test matrix for the vortex separator system.
The chiller temperature at the liquid outlet on the vortex separator will be varied depending on the temperature of water in the CO2 humidifying tank, which will affect the water temperature as shown in the above matrix. The lower the temperature of the water in the CO2 humidifying tank, the colder this chiller temperature must be set to. This is to ensure an optimal humidified stream being directed into the gas inlet of the vortex separator by the ejector. Heater temperature at the gas outlet of the separator will also be varied to ensure the purified CO2 is around room temperature, a parameter given to the X-Hab team by NASA. This value is represented by CO2 temperature. The water flow rate at the liquid outlet of the vortex separator, after the chiller, will also be varied to find a flow rate that demonstrates efficient H2O removal. This flow rate will need to be slow enough that it does not cause pressure build up at the liquid inlet nozzle and ejector, but fast enough so that it does not lose its chilled temperature.
Table 2: Test Matrix (Broken Up) for Vortex Separator
Figure 3 shows the system instrumentation and table 4 shows images and description of this instrumentation. All instrumentation is to be connected to LabView, a software that is used for data acquisition, monitoring, and recording. CO2 sensors are used with a software provided by Sensirion called “Sensirion Evaluation Kit.” Figure 4 shows the instrumentation more in depth.
Figure 1: Instrumentation of System
Table 3: Instrumentation Description
Figure 2: Overall Setup of System