How to Calibrate Scanning of TS240 using Maintenance software
How to Calibrate Scanning of TS240 using Maintenance software
Step 1 : Arrange a Clean white A4 printing paper of about 75GSM, Cut the paper neatly to approximately 11.5cm x 24cm.
Note : The paper should not be folded, crumpled,crushed,written or dirty. Make sure its clean and flat.
Step 2 : Download Maintenance software -> Download
Step 3 : Decompress the .zip file, Connect the TS240 and Open TS240_Maintanence.exe
Step 4 : The software will automatically gets connected to TS240.
Step 5 : Press Front button to calibrate the Front CIS Scanning.
Step 6 : Insert the white paper mentioned in Step 1, and press START CALIBRATION button.
NOTE : Check the "Optimize UV (NOT dll backward compatible) only if there is issue with UV images.
Step 7 : It may take few minutes to Calibrate. Don't interrupt while the process is ongoing. Once completed, "---CALIBRATION COMPLETED---" shows up at the status, Then press Exit button.
Step 8 : Press Rear button to calibrate the Rear CIS Scanning.
Step 9 : Insert the white paper mentioned in Step 1, and press START CALIBRATION button.
Step 10 : It may take few minutes to Calibrate. Don't interrupt while the process is ongoing. Once completed, "---CALIBRATION COMPLETED---" shows up at the status, Then press Exit button.
Step 11 : Disconnect and Reconnect the TS240 to complete the process Successfully.
NOTE : IN CASE OF ANY ISSUES, RESTART FROM STEP 1.