Metal Analysis
![](https://www.google.com/images/icons/product/drive-32.png)
Sample Preparation
o Acid digestion/dilution (typically HNO3)
§ Can cause proteins to crash out and interfere with nebulizer
o Alkaline digestion + chelating agent (EDTA) + Surfactant (Triton-X100)
§ Proteins more tolerant in alkaline solution
§ EDTA added help metal solubility/stability
§ Triton-X100 added to disperse and solubilize proteins
o Total Dissolved Solids (TDS) content of sample should be < 0.2%
§ Typically dilute samples between 10 and 50 times
o Internal standards
§ Should be similar in mass and ionization potential to target analyte
§ Common standards, not found in biological samples
· Lithium (6Li)
· Scandium (45Sc)
· Germanium (72Ge)
· Yttrium (89Y)
· Rhodium (103Rh)
· Indium (115In)
· Tellurium (125Te)
· Terbium (159Tb)
· Rhenium (185Re)
· Iridium (191Ir)
· Bismuth (209Bi)
o Useful for high mass analytes (i.e. Pb)
o Bismuth still used in some medications, take into consideration when designing experiments
Sample introduction
o Infused via peristaltic pump into a concentric pneumatic nebulizer
Interference
o Spectroscopic: non-analyte ions with the same m/z as target
§ Isobaric elements: isotopes of two elements that have the same m/z within the resolution of the MS
§ Double charged ions: occurs when the second ionization potential of the analyte is less than that of Argon
· Rarely occurs, but important to consider in specific circumstances (i.e. selenium and gadolinium in biological samples exposed to MRI contrast solution)
§ Polyatomic ions: form due to recombination reactions or incomplete atomization
· Highly problematic, frequently occurs due to reactions with sample matrix, reagents used in prep, plasma gasses, and atmospheric gasses (i.e. 40Ar35Cl and 75As)
· Can monitor degree of oxide formation through tuning with cerium (<3% considered acceptable)
· Careful selection of sample treatment reagents required to avoid this interference (i.e. HCl can interfere with 23V and 33As)
§ Tailing interference: due to spectral overlap from an adjacent mass
· i.e. 55Mn can experience overlap from 54Fe and 56Fe when the concentration of Fe is much greater than that of Mn in a given sample
o Non-spectroscopic: due to sample matrix or instrument drift
§ Matrix effects: enhancement or suppression due to contents of sample matrix
§ Sample introduction effects: Overloading the plasma with aerosol can cause cooling and decreased ionization.
· Can often be mitigated through further sample dilution
§ Plasma effects: enhancement or suppression of signal due to other ions present in the sample
· Suppression: High levels of easily ionizable elements (i.e. K and Na) will suppress ionization of the target analyte
· Enhancement: Presence of carbon in a sample can enhance the signal of analytes with higher ionization potentials
§ Space-charge: Most significant matrix effect, caused by electrostatic repulsion in the plasma beam.
· Lower mass analytes are more greatly affected by this phenomenon; Important to choose an internal standard with a similar m/z to the analyte of interest
o Mitigation by dynamic reaction cell (DRC): use of a reactive gas (i.e. O2) in a reaction cell prior to analysis in quadrupole
§ On mass detection: Interfering ion reacts with gas to form a polyatomic ion with a different m/z from target analyte
§ Off mass detection: Analyte reacts with gas to form a polyatomic ion with a different m/z from interfering ion
§ Neutralization: Interfering ion is neutralized by the reactive gas
Section detailing elements we are currently set up to run
o Will need to be updated as we complete new projects