a. TAs will provide a starting volume of t-pentanol (ROH)
between 0.80-1.20 mL ROH
b. Use that value to calculate the theoretical mass of product, t-pentyl chloride (RCl)
determine limiting reagent (reagent with fewest moles present in solution at start of reaction)
convert moles of limiting reagent to moles of product using molar ratio from the balanced equation
Lots of mix, vent, repeat steps! Good to make sure the septum for the vial lid won't fail.
a. Use Pasteur pipets and a pluringe to transfer reagents directly into conical vial
careful not to cross-contaminate!
b. Cap the vial and let the mixture stand for a minute
(Ex. if volume measured = 4 mL and ILE = 1 mL, reported value 4 ± 1 mL)
Instrument limit of error (ILE) is equal to the least count of the measuring device, or a fraction thereof (ILE = least count x fraction)
least count = volume represented by smallest graduation marks
fraction = 1 if the smallest graduation marks are close together
fraction = 1/2 if you can estimate half-way between the smallest graduation marks (visually)
Why are the ILE measurements (for equipment without digital displays) dependent on the user?
a. Carefully shake the mixture with occasional venting in the fume hood.
The mixture should not leak but be careful not to over-tighten the cap. As pressure builds, it will be harder to open.
b. Allow 3-4 minutes for the two phases to completely separate.
The density of t-Pentyl chloride is 0.866 g/mL at room temp. Which layer is aqueous?