Sample Information

  • Non-conductive samples must be coated by gold/platinum for high resolution images. For sputter coating, charges will be collected as mentioned in the “Usage charges” tab.

  • For solid samples the dimension should not be more than 20 mm x 20 mm x 15 mm. For any other kind of special samples which exceed the dimension mentioned above, special permission has to be sought from the instrument in-charge.

  • Solid samples must be thoroughly cleaned/ultrasonicated to prevent any contamination.

  • In case of magnetic samples, the sample condition should be informed to the operator before loading the sample for FESEM analysis.

  • Powder samples and magnetic samples must be properly reported in the requisition form.

  • Biological samples and liquid samples are strictly prohibited for FESEM analysis.