Appendix B(3) Specifications for serial numbers from 141 and up and meters having a serial number followed by a U
· Three resistance measurement ranges:1.9999 Kohms, 19.999 Ohms and 1999.9 milliOhms.
· IP (Quadrature Component) measurement range:1.9999 and 199.99 ohms
· Transmitter constant current choices:2.0 milliamps when on the 1999.9 ohm range or 20.0 milliamps when on the 19.999 ohm range or the 1.9999 ohm range
· Input resistance: 200 Megohms
· Power source: 4 standard “D” size alkaline batteries
· Accuracy: Better than 0.1 percent plus or minus 3 least significant digits (LSD) over 0°C to 50°C ambient.
· Switching frequency: 2.50 Hertz plus or minus 0.01 percent.
· No mechanical relays
· Designed to operate with stainless steel electrodes. No nonpolarizing electrodes required to operate in either the resistivity or the IP mode.
· Transmitter maximum compliance voltage: up to 640 peak-to-peak at 2 mA or 450 peak-to-peak at 20 mA (values for compliance voltage assume fresh batteries at room temperature)
· Operating Temperature Range: -10 to +50 degrees Celsius (limited by LCD)
· Storage Temperature Range: -20 to + 75 degrees Celsius
· Housing: water tight, high-impact ABS plastic
· Display device: 4.5 digit LCD Digital Panel Meter
· Analog to digital converter: laboratory grade +/- 20,000 dual slope integrated
· Method of demodulation and detection: Synchronous Demodulation
· Dimensions: 36.3 x 27.4 x 15.8 cm (14.3 x 10.8 x 6.2 inches)
· Weight: 4.3 Kilos (9.5 pounds)
· Impact resistance: will sustain a drop of one meter to a concrete surface (vibration isolated printed circuit board)
· Resistance (in phase or real) mode resolution: 0.1 milliohm
· IP (quadrature component) mode resolution: 0.01 milliohm
· Typical resistance mode capabilities: Provides accurate Wenner Array resistivities with “a” spacings of 1 foot to 1000 feet (current electrode separation of 1,800 feet or more).
· Typical induced polarization mode capabilities: Provides accurate Wenner Array IP phase readings with “a” spacings of 1 foot to over 200 feet (300 feet between current electrodes).
· Readings per set of batteries at room temperature: 1999.9 W range – 1,400 readings
19.999 W range – 750 readings