1. Full Scan MS
Q1 (quadrupole) remains in wide-pass mode.
All ions enter the TOF analyzer without precursor selection.
Provides high-resolution, accurate-mass MS1 spectra.
1.2 Typical Applications
Unknown compound discovery
Accurate mass measurement / formula prediction
PFAS initial screening
Chemical profiling of environmental samples
✔ Comprehensive, unbiased data
✔ Excellent mass accuracy
✘ No MS/MS → limited structural information
2. Targeted MS/MS
Q1 isolates a specific precursor ion (e.g., m/z 413 for PFOA).
Collision cell fragments the precursor at selected CE.
TOF records high-resolution product-ion spectra.
2.2 Applications
Confirming identity of known analytes
Targeted PFAS analysis
Drug/metabolite verification
Quantitative confirmation workflows
✔ Strong diagnostic fragments for confirmation
✔ Highest selectivity for known targets
✘ Not suitable for unknown compounds
3. Auto MS/MS (DDA: Data-Dependent Acquisition)
Perform a full MS scan.
Software automatically selects “Top N” most intense ions.
Trigger MS/MS on each selected precursor.
Apply dynamic exclusion to avoid repeated fragmentation of the same ion.
Structural elucidation of unknown PFAS
Metabolite discovery
Environmental unknown screening
Complex mixture analysis
✔ Automated, discovery-oriented
✔ Provides MS/MS for many compounds
✘ High-flow separations may miss fast eluting peaks
✘ Can generate redundant or biased fragmentation