Scheme of the HeDrop setup
Absorption and evaporation cycle of the doped droplets
The HeDrop setup has 3 major events that need to be synchronized:
Ion isolation and accumulation
The ion of interest is generated by nanoelectrospray ionization (nanoESI), their trajectory focused by ion-guide lenses and then accumulated in a hexapole ion trap.
He droplet generation and ion pickup
Superfluid He droplets of about 10000-20000 atoms are generated by an Even-Lavie valve in-axis with the ion trap. Pre-cooled ions (~90 K) are picked up by the droplets and rapidly cooled down to the droplet equilibrium temperature (~0.4 K). Due to the high kinetic energy, the doped droplet with the ion of interest can now overcome the trapping potential of the trap and exit.
Doped droplets and FEL interaction
Doped droplets exiting the ion trap reach the so-called interaction region in which they are hit by the IR light of the FEL. If the photon energy matches a vibrational transition, radiation is absorbed by the ion and rapidly dissipated to the helium droplet (tens of fs), resulting in the evaporation of helium atoms. The FEL radiation consists of macro-pulses of ~10 μs, with micro-pulses of 1-10 ps separated by 1 ns (see FHI-FEL). The absorption-dissipation event in the droplet occurs after the arrival of every FEL micro-pulse until complete evaporation of the helium atoms. The bare ion is then monitored by a time-of-flight analyzer and its intensity plot against the photon energy.