EDX uses the characteristic X-rays generated by electron–specimen interactions to identify the elements in a material.
The electron beam strikes the specimen and can eject electrons from inner atomic shells.
When an electron from a higher-energy shell fills the vacancy, the atom may release an X-ray with an energy characteristic of that element and electronic transition.
The EDX detector measures the energies and intensities of the incoming X-rays.
The instrument generates an EDX spectrum displaying X-ray energy, usually in kiloelectronvolts (keV), against signal intensity or counts.
Different peaks correspond to characteristic X-ray lines associated with particular elements.
The spectrum is interpreted to identify the elements detected in the analysed region.
Depending on the analytical conditions and data-processing method, the results may include qualitative identification and estimated elemental composition.
What can EDX reveal?
Elements present in a specimen
Relative elemental distribution
Approximate elemental composition
Elemental differences between selected regions
Spatial distribution of selected elements through elemental mapping
Important note: EDX generally does not identify chemical bonding or determine a material's oxidation state by itself. Its ability to detect trace or low-concentration elements also depends on the specimen, instrument settings, peak overlaps and other analytical limitations.
Suggested visual: An EDX spectrum labelled with Energy (keV) on the horizontal axis and Counts on the vertical axis, accompanied by an example elemental map.