EVD Manipulation
EVD Manipulation Protocol:
EVD Manipulation Protocol:
- All staff in room wear cap and mask
- 3-way stopcock of access port turned to 45 degrees (OFF to all directions)
- Access port and surrounding tubing submerged in isopropyl alcohol [panel C]
- Physician dons cap, mask, sterile gown, and sterile gloves
- Access port and surrounding tubing cleaned with chlorhexidine [panel D]
- Port cap removed and discarded
- Inside of port cleaned with chlorhexidine multiple times [panels E-G]
- Inside of port rinsed with sterile, preservative-free NS multiple times [panel H]
- CSF draw or flush performed with strict sterile technique
- New sterile port cap applied
This work is licensed under a Creative Commons