Search this site
Embedded Files
Skip to main content
Skip to navigation
Home
Contact
Home
Contact
More
Home
Contact
No flow
Liver on board
We suggest the following remedial steps:
Check that metra is in 'liver on-board' mode
Confirm pump is running (check GUI screen)
Re-wet flow probes
Check tubing for kinks
Check white Roberts clamps in liver bowl (should be fully open)
Confirm PVHa is not at 100% occluded by checking top left of GUI screen - stop perfusion (for four seconds) and restart to reset PVHa
Stop perfusion, wait 5 seconds and restart
Power down and restart metra
If problem persists, please
contact us
.
Google Sites
Report abuse
Google Sites
Report abuse