Make&Model: Agilent Technologies/ 5800 ICP-OES (Trace element detection tool)
Specification:
The ICP-OES spectrometer: To determine the elements in diverse kinds of samples
RF Generator: solid state Free Running with 40 or 27 MHz frequency
Power output stability: 1000 - 1500 Watts 1 watt increments in both Radial for, Axial, and dual view analysis with auto tuning.
Power efficiency of RF generator: be greater than 79% with < 0.1% variation in output power stability.
Plasma flow gas consumption: ICP-OES instrument should be 8L/min or less for an aqueous sample.
Nebulizer argon gas flow: variable between 0 and 2.0 L/min in 0.01 L/min increments.
OPTICS: dual monochromator/polychromator based Spectrometer with Echelle Grating and Optical Resolution of minimum 0.009 nm or better at 200 nm.
Multielement Standards: Ar, Be, Cd, Cr, Pb, Hg, Ni, Se, Tl, Al, Ba, Bi, B, Ca,Co, Cu, Ga, In, Fe, Li, Mg, Mn, K, Ag, Na, Sr, Zn
Sample Type: Aqueous or solid sample can be analyzed.
Microwave Digester: Solid samples can be digested.
Application: multiple metals can be detected in a sample.
Minimum Volume: 10 ml acid mixture is required for digestion of a solid sample.