Medical Compact Backscatter Imaging

Designing and prototyping a compact handheld x-ray 

Spring 2023 MAE 156B Senior Design Project

University of California, San Diego

BACKGROUND

Current X-ray imaging devices are based on “transmission mode imaging” which is a process where photons are sent through an object and are detected on the other side in order to produce an image. This practice requires high energy photons in order for the photons to completely pass through the object.  

However an alternative imaging method, backscatter imaging, instead acquires an image by detecting the photons that reflect back from the object. Backscatter imaging utilizes lower energy photons, and therefore is less dangerous in terms of human radiation exposure, as well as in terms of high voltage through the device, as less energy is required to generate an image. This technique is commonly used in a variety of engineering and R&D applications, but it is not currently used in medicine as acquired images are not as clear as traditional imaging. The objective of this project is to not only design such a device, but also to determine the photon energy that yields an optimal image quality. Such a device has potential for many different medical situations, particularly for spinal or breast cancer applications.  

WHY THIS IS IMPORTANT / INSPIRATION

Traditional transmission x-ray is not optimal for the imaging of shallow tissues as photons must penetrate through the body of a patient in order to be detected. This results in the introduction of extraneous information from tissues and structures behind the target, distracting from or obscuring important elements in the radiological image. Furthermore, penetration throughout the entire body requires higher energy photons as photons must be capable of passing through more material, as well as more overall photons, as photons may scatter away from the detector when they contact patient tissue and bones. This has the potential to expose patients to more radiation than may be necessary and to cause health issues, such as cancer, down the line for the patient.

  In contrast, a backscatter imaging would only require direct access to the tissues being imaged, as the detector and device would be on the same side, resulting in an increased access to and ease of radiological imaging in situations where patient mobility is restricted, such as an operating room. 

Additionally, such a device would generate an image using fewer overall photons, as photons would not have to penetrate throughout the entire body structures of the patient, reducing the number of opportunities for photons to be absorbed by tissue or otherwise scattered away from the detector. Furthermore, as photons do not need to be able to penetrate through large amounts of dense objects (such as bone), the photons required for imaging are not required to carry large amounts of energy as is the case with transmission imaging. Therefore, imaging through the backscatter modality has the potential to expose patients to significantly lower amounts of radiation. 

OBJECTIVE

The first objective is to find the optimal X-ray energy by utilizing a Monte Carlo simulation of breast tissue with calcifications and titanium marker clips.

The second objective is to design and prototype an X-ray source with the associated electronic components. The designing of the photon sources must be able to detect a 1 cm3 object of titanium and calcium under a skin. The x-ray sources must be no larger than 512 cm3.

Simulation Results

A simulation is required to determine the optimal energy of a backscatter imaging device due to safety considerations. Given the complex interactions between photons and the medium, a Monte Carlo simulation is appropriate for this project.  Varying energy levels of photons can results to a different outcome.

The animation shows the simulation process where the source travers through the tissue. The photons exhibits a distinct  behavior when encountering the titanium. 

The graphs show the number of photons detected for backscatter indicate that the optimal energy range is between 30 -40 keV.

FINAL DESIGN OVERVIEW

Our final design consists of a housing unit for the x-ray source (x-ray tube) which has a built in scanning mechanism


Performance Results