X-ray Diffraction Data Collection
Roger S. Rowlett
Gordon & Dorothy Kline Professor, Emeritus
Colgate University Department of Chemistry
Gordon & Dorothy Kline Professor, Emeritus
Colgate University Department of Chemistry
Before a data set can be collected—from which the final structure of the protein can be deduced—it is necessary to mount a crystal in the X-ray beam of a diffractometer and determine if it diffracts to sufficient resolution to justify collecting a full data set. The initial crystal screen can often be used to determine the space group of the crystal, an important piece of information is planning data collection. The exact instructions for collecting data will vary depending on the type of equipment used. Instructions for several facilities are given here.
The following instructions are for collecting protein data using a Gemini R instrument located in the Department of Chemistry (Wynn Hall) on the Colgate University campus.
Figure 1. The Colgate University Department of Chemistry Oxford Diffraction Gemini R system.
Note: Current recommended settings for the cyrojet are 6.5 L/min sample and 4.5 L/min shield flow for 110 K and good frost suppresssion. At this setting, the cryojet will consume about 40% of the cryojet dewar capacity per day of operation. It is programmed to auto-fill at 25% capacity to a final fill of 90%. Therefore, it should be expected to autofill about every 36-40 hours. A full 160 L dewar of LN2 can be reliably expected to perform two full autofills. Do not start an experiment unless there is sufficient LN2 to complete the run. Long runs (more than 2 days) may require attachment of a new 160 L LN2 dewar to the autofill. Helium consumption for the Cu source will deplete about 100 psi per day of operation. Do not start an experiment unless there is sufficient helium pressure to complete the run.
The default settings for the autofill are
This settings will result in an autofill about every 36 hours during normal cryojet operation. I may be desirable to change these settings slightly to maintain cryojet temperature stability, and schedule autofills more frequently during data collections. The following settings will schedule an autofill when the LN2 level drops to 50% in the cryojet dewar, and will result in an autofill approximately every 24 hours so the filling can be easily monitored during working hours:
When the supply LN2 dewar is low (i.e., insufficient LN2 to complete a full fill when the next autofill is triggered) it can be advantageous to perform a manual fill of the 75L cryojet dewar. Use the following steps to do a manual fill:
These instructions assume you will be using a magnetic goniometer head that is suitable for Hampton Research 18 mm crystal caps. Our instrument has both Hampton Research and Huber goniometer heads. The Huber is normally installed on the goniometer head.
Note: The arguments for the "goto angle" command are in the order omega, theta, phi, kappa, and distance.
This method will reset the goniometer to vertical position before collecting screening scans. The goniometer will return to the previously set position (usually the mounting geometry with OTKPD = 40 -30 80 -90 120) when the crystal centering dialog is closed. This can be time consuming if many crystals are screened, but the mounting geometry is very easy.
This method will keep the goniometer in the LOWER crystal centering position during data collection, so that the goniometer geometry does not have to switch between the mounting geometry, the screening geometry, and the crystal centering geometry for each run. Mounting crystals on the goniometer in the LOWER crystal centering position is a little more difficult, but quite possible.
Distance (mm), R(edge), R(corner)
60, 2.26, 1.92
65, 2.39, 2.01
70, 2.52, 2.11
75, 2.66, 2.20
80. 2.79, 2.30
85, 2.93, 2.40
90, 3.07, 2.50
95, 3.21, 2.61
100, 3.35, 2.71
105, 3.49, 2.81
110, 3.64, 2.92
115, 3.78, 3.03
120, 3.92, 3.13
Note: the "edge" of the 135 mm ruby detector is about 100 mm when processed in CrysalisPro. The "corner" is actually the edge of the circular detector, about 135 mm; it does not extend to the corner of the square display area in CrysalisPro.
Collecting data with CrysalisPro is highly automated, and this mode of operation is encouraged. Collecting data typically involves collecting a pre-experiment to establish appropriate values for signal to noise ratio and resolution, selecting a strategy to accomplish the collection of adequate data in the time available, and then initiating the full experiment.
The pre-experiment will collect several frames of data at each of three omega angles to learn about the scattering properties and space group of your crystal, and provide guidance for selecting a suitable collection strategy.
Note: The I/σ value in CryslalisPro refers to unmerged data without accounting for redundancy. The final value of I/σ for the scaled and merged data set can be expected to be approximately equal to the entered I/σ value times the square root of the redundancy. For example, an experiment with a target I/σ of 6 and a target redundancy of 4 will result in a scaled and merged I/σ value of approximately 12.
It is typically necessary to alter the initial suggested strategy before starting the experiment. Many options are available to configure the final experiment.
Note: If planning to process data in MOSFLM, ensure that all strategy runs are of the "omega scan" type.
Data collection will start immediately after clicking on the Start experiment button in the strategy module. Progress of data collection can be followed by examining the information cards on the right of the CrysalisPro screen.
CrysalisPro will re-analyze the data after every 25 frames, and at the completion of each set of strategy runs
Perform the following operations when data collection is complete. If another user will follow you and require the cryojet, you may omit the cyrojet shutdown procedure.