Excessive difference in temperature between the heater unit and the perfusate occurs primarily when the Terumo is not able to read temperature correctly. Please check the following:
Ensure white filter is removed from below shunt sensor
Ensure that water heater tubing is correctly fitted. Connectors should click when pushed firmly into place.
Ensure both white Roberts clamps are fully released. Check that the line is filled with blood and is warm to touch.