Home‎ > ‎Instruments‎ > ‎

Rigaku Small-angle X-ray Scattering System XRD

Rigaku Small-angle X-ray Scattering System


Types of samples

Polymers, biomaterials, fibers, colloids, metals, and liquid crystals
Information obtained: Domain size, shape, and distribution.
Information is statistically averaged and must be modeled.


Contact:
Donald Savage
dsavage@facstaff.wisc.edu
608-263-0831








Instrument specifications

X-ray source – micro focus Cu Kα sealed tube

With cross coupled multilayer mirrors

• operates at 40 W
• 20 µm by 20 µm source size

Automatic multiple sample scanning capability
Heating and cooling: Linkam temperature stage (single sample only) allows -50C to 300C with 0.5C degree precision. A liquid cell is available.
Sample to detector length: 2 meters
Detector is 2-dimensional multi-wire
with 1024 by 1024 wires over a 6-inch area.

In highest resolution configuration:

Minimum Angle measured – 5.8 x 10 –4 radians
• qmin (Angstroms –1) = 0.0066
• dmax (Angstroms) = 950
• beam size at detector < 1.5 mm