Panametrics 25DL HP

Quick setup Guide

Press

On/Off Key

The unit will turn on

Select Which transducer you are going to use.

Three are supplied with the kit.

  1. M102 General Purpose
  2. M106 Thick Composites and Fibreglass
  3. M2008 Thiner Composites and Fibreglass

You may need to experiment with each probe to determine which is the most appropriate.

To configure the unit you will need two test pieces of the same material which you want to test that each have known thicknesses. One thick and one thin.

 

Press

2nd F

 

Press

CAL (Setup)

A list of transducers will appear

Use the Up and Down arrow keys and select the transducer which you have decided to use

Press

Enter

 

Press

MEAS

 

  • If you are using the M2008 transducer hold the transducer in the air and press Zero briefly
  • Place couplant on the thick material.
  • Hold the transducer to the thick test piece
  • Wait for a stable reading

Press

CAL(Setup)

If locked appears press MEAS then 6&3 together to unlock

Press

VEL

CAL & VEL will appear on the display

  • Enter the known thickness of the thick test piece using the keypad
  • Place couplant on the thin material.
  • Hold the transducer to the thin test piece
  • Wait for a stable reading

Press

CAL(Setup)

 

Press

ZERO

CAL & ZERO will appear on the display

  • Enter the known thickness of the thick test piece using the keypad

Press

MEAS

This will save the calibration figure

More detail of this procedure is shown in section 3.3 of the manual.