4~20mA Outputs Notes
Range of Output 1 |
|
· Field selectable from the
following:
|
|
Output Selection
|
Range
|
|
Linear oxygen
|
0 – 0.1% oxygen to 0 – 100.0% oxygen
0 – 1000ppm oxygen to 0 –
1,000,000ppm oxygen
|
|
Low Range Linear oxygen
|
0 – 0.001% oxygen to 0 – 1.0% oxygen
0 – 10ppm oxygen to 0 – 10,000ppm
oxygen
|
Range of Output 2
|
|
· Field
selectable from the following:
|
|
Output Selection
|
Zero Range
|
Span Range
|
|
|
Carbon Dioxide
|
0 – 90%
|
10 – 100%
|
Min span 10%
|
|
Reducing Oxygen
|
10-1 – 10-10 %
|
10-1 – 10-30 %
|
Min span three decades
|
|
In one decade steps.
|
|
Oxygen sensor EMF
|
0 – 1100mV
|
1000 – 1300mV
|
|
|
|
In 100mV steps
|
In 100mV steps
|
|
|
Sample Oxygen
|
0 – 0.1%
|
0.01 – 20%
|
Min Span 0.01%
|
|
|
0 – 1000ppm
|
100 – 20,000ppm
|
Min span 100 ppm
|
|
Low oxygen
|
0 – 99.9%
|
0.1 – 100% oxygen
|
Min span 0.1%
|
|
|
0 - 999,000ppm
|
1,000 - 1,000,000ppm
|
Min span 1,000ppm
|
|
Logarithmic Oxygen
|
0.1% O2
|
Fixed 20% O2 Fixed
|
|
4.6 4 - 20 / 0 - 20mA OUTPUTS
There are two analog output channels.
The output function and range can be
selected for both channels. See set-up 25 to 30.
NOTE: If oxygen is selected to be transmitted on either channel,
the output current will be driven to 20mA while the sensor
heats up to over 650°C.
NOTE: If “Not Used” is selected for channel 1 output, menu items
that relate to both output channels will be hidden.
14. OUTPUT CHANNEL #1, 0 - 20mA or 4 - 20mA
The output channel can be scaled 0 to 20mA
or 4 to 20mA to represent the parameter that is selected in set-up function
25.
Options:
1. Not Used *
2. 4 - 20mA
3. 0 - 20mA
15. OUTPUT CHANNEL #2, 0 - 20mA or 4 - 20mA §
The output channel can be scaled 0 to 20mA
or 4 to 20mA to represent the parameter that is selected in set-up function
28.
Options:
1. 4 - 20mA *
2. 0 - 20mA
16. 4 to 20mA CALIBRATION OPTIONS, CHANNEL #1 §
Select the calibration method for the 4 -
20mA output channel #1.
The output channels can be either
calibrated by simply pressing the ‘AUTO
CAL’ button, or can be trimmed at
both the
4mA and 20mA ends of the scale using an
external multimeter.
Options:
- Auto Calibration *
- Manual Calibration
- Set 4mA Trim
- Set 20mA Trim
If ‘AUTO CAL’ is selected, the output channel is calibrated when ‘Auto Cal’
is initiated from the keyboard (See section
6.3).
If ‘MAN
CAL’ is selected, it is necessary
to trim both ends of the 4-20mA output range using the 4mA and 20mA
options in this menu item. Selecting ‘MAN CAL’ inhibits the ‘Auto Cal’
process of this channel.
Always do the 4mA trim first, and then the
20mA trim. After trimming both ends of the scale, return the
‘CALIBRATION OPTIONS’ menu option back to ‘MAN CAL’ (not ‘AUTO CAL’), or the
calibration factors will be over written by the next ‘AUTO CAL’.
For more details on calibrating the output
channels, see section 6.3.
NOTE: The analyser will only stay in
either ‘4mA TRIM’ or ‘20mA TRIM’ modes for 30 minutes before it
automatically returns to ‘MANCAL’.
17. CALIBRATE 4mA, CHANNEL #1 §
This menu item only appears if ‘Set 4mA
Trim’ is selected in Set-up 16.
Range: 0 to 25mA, Default is 4.00mA
For full details on the calibration of the
4 - 20mA output channels, see section 6.3.
18. CALIBRATE 20mA, CHANNEL #1 §
This menu item only appears if ‘Set 20mA
Trim’ is selected in Set-up 16.
Range: 0 to 25mA, Default is 20.00mA
19. 4 to 20mA CALIBRATION OPTIONS, CHANNEL #2 §
Select the calibration method for the 4 -
20mA output channel #1.
For more details, see Set-up 16 and
section 4.5.
Options:
- Auto Calibration *
- Manual Calibration
- Set 4mA Trim
- Set 20mA Trim
20. CALIBRATE 4mA, CHANNEL #2 §
This menu item only appears if ‘Set 4mA
Trim’ is selected in Set-up 19.
Range: 0 to 25mA, Default is 4.00mA
For full details on the calibration of the
4-20mA output channels, see section 6.3.
21. CALIBRATE 20mA, CHANNEL #2 §
This menu item only appears if ‘Set 20mA
Trim’ is selected in Set-up 19.
Range: 0 to 25mA, Default is 20.00mA
6.3 D/A (4 - 20mA output channels) Calibration
The calibration can either be done using
the ‘Auto Cal’
or ‘Manual Cal’.
Auto Cal
The ‘Auto
Cal’ mode is selected in set-up
16 (and 19 for channel 2).
The analyser will automatically divert the
output back to the input, measure the offset and span, and record the calibration
factors for each channel.
If either of the channels are selected to
be calibrated manually, the factors will not be changed by an ‘Auto Cal’. Manual Cal
The ‘Manual Cal’ mode is selected in
set-up 16 (and 19).
Set the 4mA calibration first and then the
20mA calibration.
- Select ‘Set 4mA Trim’ in set-up 16 (or 19).
- Return to RUN mode.
- Measure the output on the channel to be calibrated with a digital
multimeter. If the current is not exactly 4.00mA, return to set-up mode
and change the 4mA calibration factor in set-up 17 (or 20).
- Re-measure the current while back in RUN mode until the current is
within 3.9 to 4.1mA.
- Return to set-up mode and select ‘Manual Cal’ in set-up 16 (or
19).
Set the 20mA calibration factor.
- Select ‘Set 20mA Trim’ in set-up 16 (or 19).
- Return to RUN mode.
- Measure the output on the channel to be calibrated with a digital
multimeter. If the current is not exactly 20.00mA, return to set-up mode
and change the 20mA calibration factor in set-up 18 (or 21).
- Re-measure the current while back in RUN mode until the current is
within 19.9 to 20.1mA.
- Return to set-up mode and select ‘Manual Cal’ in set-up 16 (or
19).
This calibration is now saved in battery
backed memory until
- The factors are changed in the manual
calibration
- The analyser is forced into a COLD-START (see
section 6.1)
- The calibration mode in set-up 16 (or 19) is
changed to Auto Cal and an Auto Cal is initiated.
NOTE: The
4mA or the 20mA trim mode will only be held on the output channels for 30
minutes before automatically returning to ‘Manual Cal’ mode in set-up 16 (or
19).
Open the case by removing the two screws either side of the display. The connector for the analogue output can now be seen.